2.4 Measurement operation
AE-4022
2-17
2.4
Measurement operation
2.4.1
Normal measurement
Normal measurement mode is used for consecutive measurement of samples. To measure one or more urgent samples
immediately during a normal measurement process, press
or
to switch to the STAT measurement mode.
1
Prepare the samples
a
Prepare the samples, referring to “2.3.5. Sample preparation” on page 2-13.
2
Set the measurement number
After the first sample is allocated a 4-digit measurement number, the system automatically increments the
measurement number by one. These measurement numbers are stored in the system until the power is turned OFF. To
set the first measurement number to “0001”,
do not
enter a measurement number, but follow the procedure in step
below.
a
Enter a measurement number using the numeric keys and press
.
• You can enter any value from 0 to 9999. When the measurement
number “9999” is incremented, the sequence will begin again
from “0000”.
3
Set the patient ID number (when required)
a
Enter the patient ID number, referring to “2.3.6. Entering patient
ID numbers” on page 2-14.
Wear protective gloves to prevent exposure to pathogenic microorganisms.
Содержание AUTION ELEVEN AE-4022
Страница 1: ......
Страница 75: ...Chapter 2 Measurement Operation AE 4022 2 40 ...
Страница 112: ...3 7 SETUP user settings AE 4022 3 37 4 Quit the SETUP menu aPress twice The display returns to the Standby screen ...
Страница 143: ...Chapter 4 Maintenance AE 4022 4 14 5 Close the printer cover aClose the printer cover Printer cover ...
Страница 146: ...AE 4022 5 1 Chapter 5 Troubleshooting 5 1 Warning messages 5 2 5 2 Error messages 5 3 5 3 Trouble messages 5 5 ...
Страница 153: ...Chapter 5 Troubleshooting AE 4022 5 8 ...
Страница 160: ......
Страница 161: ......