2.4 Measurement operation
AE-4022
2-21
10 Prepare the next sample and test strip
a
Repeat steps
for entering the clarity value, and dipping
and setting subsequent test strips.
• When a test strip is placed, the system automatically starts the
measurement of the next sample.
11 Finish normal measurement
a
When the normal measurement finishes, the display returns to the
[Standby screen].
The measurement and patient ID numbers for the next
measurement are displayed on the [Standby screen].
REFERENCE:
The measurement number can be changed during the
measurement process. To change the measurement
number from “0005” to “0050”, for example, input
and press
.
After displaying “---”, the system changes the measurement
number to the input number and then the screen returns to
the [Measurement screen].
To interrupt measurement, press
. The system then
stops measurement and the display returns to the [Standby
screen].
The measurement and patient ID numbers for the next
measurement are displayed on the [Standby screen].