2.4 Measurement operation
AE-4022
2-19
6
Call up the [Clarity input screen]
a
Press and hold
for about 2 seconds.
• The [Clarity input screen] appears.
7
Enter the clarity value
a
Enter the clarity value of the sample using the numeric keys.
• Enter a number between 1 and 5 using the table below.
b
Press
.
• The display returns to the [Standby screen].
8
Dip a test strip in the sample for 2 seconds
a
Dip the test strip in the sample for 2 seconds, and then withdraw it.
Input No.
1
2
3
4
5
Clarity
Clear
Slightly
cloudy
Cloudy
Turbid
Other
REFERENCE:
When the clarity value of the sample is “1”, just press
without pressing any numeric keys.
IMPORTANT:
Dip the entire test strip pad area in the sample at once.
Do not
allow the
black mark shown in the figure to become wet. If the test strip is dipped too
deeply and the black mark is moistened, correct measurement may not be
obtained.
After dipping the test strip in the sample, gently remove excess urine by
drawing the entire length of the test strip against the rim of the collection cup.
In the “Auto start” mode, the buzzer does not sound for the first measurement. Always dip the test strip for 2
seconds. Insufficient dipping of test strips may cause insufficient color change, while dipping too long may
cause reagents to drain out of the test strip, either of which will prevent correct measurement.
Do not
dip the
black mark.
Содержание AUTION ELEVEN AE-4022
Страница 1: ......
Страница 75: ...Chapter 2 Measurement Operation AE 4022 2 40 ...
Страница 112: ...3 7 SETUP user settings AE 4022 3 37 4 Quit the SETUP menu aPress twice The display returns to the Standby screen ...
Страница 143: ...Chapter 4 Maintenance AE 4022 4 14 5 Close the printer cover aClose the printer cover Printer cover ...
Страница 146: ...AE 4022 5 1 Chapter 5 Troubleshooting 5 1 Warning messages 5 2 5 2 Error messages 5 3 5 3 Trouble messages 5 5 ...
Страница 153: ...Chapter 5 Troubleshooting AE 4022 5 8 ...
Страница 160: ......
Страница 161: ......