Chapter 4 Maintenance
AE-4022
4-2
4.1
Daily maintenance
At the end of each working day, turn OFF the instrument, and then clean the feeder, and the waste box.
4.1.1
Cleaning the feeder
Sample residue often builds up on the test strip tray, carrying arm, suction ports, and test strip feed mechanism, as these
parts carry test strips. After repeated measurements, significant sample residue may accumulate. Therefore, the test
strip tray, carrying arm, suction ports, and test strip feed mechanism should be cleaned on a daily basis after use.
Items required: Alcohol, cloth, and protective gloves
1
Turn OFF the instrument.
a
Check that the [Standby screen] is displayed, and then turn OFF
the instrument.
Wear protective gloves to prevent exposure to pathogenic microorganisms.
Discard used test strips in accordance with local regulations for biohazardous waste.
NOTE:
When cleaning the test strip tray, avoid the use of organic solvents such as alcohol and thinner or ultrasonic
cleaning. These cleaning methods may deform or discolor the test strip tray and thus make it unserviceable for
further testing.
Содержание AUTION ELEVEN AE-4022
Страница 1: ......
Страница 75: ...Chapter 2 Measurement Operation AE 4022 2 40 ...
Страница 112: ...3 7 SETUP user settings AE 4022 3 37 4 Quit the SETUP menu aPress twice The display returns to the Standby screen ...
Страница 143: ...Chapter 4 Maintenance AE 4022 4 14 5 Close the printer cover aClose the printer cover Printer cover ...
Страница 146: ...AE 4022 5 1 Chapter 5 Troubleshooting 5 1 Warning messages 5 2 5 2 Error messages 5 3 5 3 Trouble messages 5 5 ...
Страница 153: ...Chapter 5 Troubleshooting AE 4022 5 8 ...
Страница 160: ......
Страница 161: ......