Chapter 3 Supplementary Operations
AE-4022
3-50
3.7.23 No.021: QC lock-out function ON/OFF
Use this submenu to specify whether to use the QC lock-out function.
1
Call up the screen
a
Call up the [Item number input screen], referring to section
“3.7.1. Operation of user settings” on page 3-17, if necessary.
b
Press
.
c
Press
.
• The [Password input screen] appears.
2
Enter the password
a
Enter the password using the numeric keys.
b
Press
.
• The [QC lock-out function ON/OFF setting screen] appears.
NOTE:
When the QC lock-out function is ON, only the control measurement mode can be selected once the QC
deadline has elapsed. The normal and check measurement modes can be selected once control
measurement has been performed.
REFERENCE:
When the QC lock-out function is set to Prompt, normal measurements can be performed even after the QC
deadline has elapsed, but “COM:W008” is printed in the measurement results report.
When the QC lock-out function is OFF, the settings for “No.020: QC deadline” are ignored.
If you press
during operation or input, the current setting is canceled and the display returns to the [Item
number input screen].
* * * *
The actual password appears on the display
as it is entered; it is not hidden by asterisks.
Содержание AUTION ELEVEN AE-4022
Страница 1: ......
Страница 75: ...Chapter 2 Measurement Operation AE 4022 2 40 ...
Страница 112: ...3 7 SETUP user settings AE 4022 3 37 4 Quit the SETUP menu aPress twice The display returns to the Standby screen ...
Страница 143: ...Chapter 4 Maintenance AE 4022 4 14 5 Close the printer cover aClose the printer cover Printer cover ...
Страница 146: ...AE 4022 5 1 Chapter 5 Troubleshooting 5 1 Warning messages 5 2 5 2 Error messages 5 3 5 3 Trouble messages 5 5 ...
Страница 153: ...Chapter 5 Troubleshooting AE 4022 5 8 ...
Страница 160: ......
Страница 161: ......