Chapter 4 Maintenance
AE-4022
4-8
13 Attach the suction ports to the test strip tray
a
Push the suction ports straight into the test strip tray until they
come in contact with the bottom of the tray (until the protrusions
on both sides are no longer visible).
14 Mount the test strip tray
a
Align the tabs of the test strip tray with the holes of the unit as
shown in the right figure, and place the tray inside the unit. Then
slide the test strip tray further inside.
Insert it until you hear a click.
15 Clean the white plate
a
Wipe any dirt off the white plate, using a clean cloth.
NOTE:
If the suction ports are not correctly in contact with the
bottom of the test strip tray, the test strip may not be
transported properly, leading to a strip jam or trouble.
White plate