1.1 Outline of the AE-4022
AE-4022
1-3
Abnormal color detection
The instrument has an abnormal color detection function that detects abnormal color generation on the reagent pad.
The instrument prints a “!” mark with the measurement result (only applicable to measurement items KET, BIL, and
URO).
Operator ID function
This function restricts instrument use by requiring operators to enter an ID number to operate the analyzer. This can
lock out persons without an ID number from using the instrument.
Either of the following authorities or both can be assigned to individual operators (or ID numbers) for the restrictions
on setting operations and measurement:
• Access authority to the SETUP menu (user settings)
• Measurement authority to override QC lockout if needed to continue testing and run QC at a later time (when QC
lock-out function is ON).
QC lock-out function
This function locks the instrument from performing measurements when QC has never been performed since the
instrument was installed or the deadline set by the operator has elapsed. The lock is released once control
measurements are performed.
* When the operator ID function is ON, the access authority to the SETUP menu is required to set up the QC lock-out function.
Clarity input function
This function allows the operator to manually input the value of urine clarity obtained visually in order to add it to the
measurement results report. (This instrument
does not
have a function to automatically determine urine clarity.)
* When the Operator ID function is ON, the access authority to the SETUP menu is required to set up the clarity input function.
Prints measurement results in bold
The instrument prints measurement results with other data emphasizing the results by using bold-type and larger
characters for easy and fast reading.
(See “2.5. How to read the measurement results” on page 2-37)
Easy to maintain
Components that require daily maintenance, such as the carrying arm, test strip tray, and waste box, have simple
structures that can easily be detached and reattached.
Store up to 520 measurement results
The instrument can store up to 520 measurement results. When the number of measurement results exceeds 520, the
instrument automatically deletes the oldest sample data. The measurement results are stored in categories of
measurement types and result types (normal and abnormal).
Optional hand-held barcode reader
A hand-held barcode reader (optional) can be mounted. Each barcode scanned by the barcode reader is allocated a
patient ID number.
Ethernet connection
An Ethernet unit (optional) can be installed.
Содержание AUTION ELEVEN AE-4022
Страница 1: ......
Страница 75: ...Chapter 2 Measurement Operation AE 4022 2 40 ...
Страница 112: ...3 7 SETUP user settings AE 4022 3 37 4 Quit the SETUP menu aPress twice The display returns to the Standby screen ...
Страница 143: ...Chapter 4 Maintenance AE 4022 4 14 5 Close the printer cover aClose the printer cover Printer cover ...
Страница 146: ...AE 4022 5 1 Chapter 5 Troubleshooting 5 1 Warning messages 5 2 5 2 Error messages 5 3 5 3 Trouble messages 5 5 ...
Страница 153: ...Chapter 5 Troubleshooting AE 4022 5 8 ...
Страница 160: ......
Страница 161: ......