2.4 Measurement operation
AE-4022
2-23
3
Set the measurement number
After the first sample is allocated a 4-digit measurement number, the system automatically increments the
measurement number by one. The measurement numbers are stored in the system until the power is turned OFF. To
set the first measurement number to “0001”,
do not
enter a measurement number, but follow the procedure in step
below.
a
Enter a measurement number using the numeric keys and press
.
• Your can enter any value from 0 to 9999. When the measurement
number “9999” is incremented, the sequence will begin again
from “0000”.
4
Set the patient ID number (when required)
a
Enter the patient ID number, referring to “2.3.6. Entering patient
ID numbers” on page 2-14.
5
Prepare test strips
a
Take out the required number of test strips from the bottle.
b
Close the test strip bottle cap securely.
REFERENCE:
The currently set test strip type is displayed on the [Standby
screen]. To change the test strip type setting, see “3.7.5.
No.002: Test strip type” on page 3-22.
IMPORTANT:
After taking the test strips out of the bottle, immediately
close the cap, or the test strips in the bottle will absorb
moisture or dirt in the air, which may deteriorate reagents
and make the test strips useless.
Содержание AUTION ELEVEN AE-4022
Страница 1: ......
Страница 75: ...Chapter 2 Measurement Operation AE 4022 2 40 ...
Страница 112: ...3 7 SETUP user settings AE 4022 3 37 4 Quit the SETUP menu aPress twice The display returns to the Standby screen ...
Страница 143: ...Chapter 4 Maintenance AE 4022 4 14 5 Close the printer cover aClose the printer cover Printer cover ...
Страница 146: ...AE 4022 5 1 Chapter 5 Troubleshooting 5 1 Warning messages 5 2 5 2 Error messages 5 3 5 3 Trouble messages 5 5 ...
Страница 153: ...Chapter 5 Troubleshooting AE 4022 5 8 ...
Страница 160: ......
Страница 161: ......