Chapter 5 Troubleshooting
AE-4022
5-6
Message
Description
Possible cause
Measures
XX XXXX XXXX XXXX
Unknown trouble
Trouble occurred due to unknown
cause(s). (The message varies
according to the trouble that has
occurred.)
Note the displayed details, turn OFF
the instrument, and contact your
local distributor.
T101
EEPROM trouble
The EEPROM is malfunctioning.
Turn OFF the instrument and contact
your local distributor.
T102
Changed version
The ROM has been upgraded.
T110
No calibration curve
The calibration curve for the selected
test strip was not entered, or else
some test strip information was
missing.
T120
Inlet error
• There was an obstacle in the test
strip inlet.
• The motor-driven part
malfunctioned.
Turn OFF the instrument and open
the maintenance cover.If there are
any obstacles or scattered test strips
inside, remove them. Check for any
damage to the carrying arm, test strip
tray, suction ports, or test strip feed
mechanism. If any of these parts
show signs of damage or the trouble
recurs, contact your local distributor.
T121
Feeder trouble
• The test strip feed mechanism
encountered an obstacle.
• Test strips are scattered inside the
instrument.
• The motor-driven part
malfunctioned.
T123
Incoming strip sensor
trouble
• A test strip was blocked by the
suction ports and was not
transferred smoothly.
• There was an obstacle in the
sensing section (such as a misfed
test strip) that desensitized the
incoming strip sensor.
• The incoming strip sensor
malfunctioned.
• Turn OFF the instrument and open
the maintenance cover. If there are
any obstacles or scattered test
strips inside, remove them. Check
for any damage to the carrying arm,
test strip tray, suction ports, or test
strip feed mechanism. If any of
these parts show signs of damage
or the trouble recurs, contact your
local distributor.
• Check if any urine has adhered to
the feeder, and clean it if
necessary.
• If the waste box is full of used test
strips, empty the box.
T130
Photometric section
initialization trouble
• The photometric section driving
mechanism, or the position
detection sensor of the photometric
section driving mechanism
malfunctioned.
• The waste box was completely full
of test strips, which prevented
proper operation of the photometric
section.
• Turn OFF the instrument and open
the maintenance cover. If there are
any obstacles or scattered test
strips inside, remove them. Check
for any damage to the photometric
section. If the photometric section is
damaged, or the trouble recurs with
no identifiable cause, contact your
local distributor.
• If the waste box is full of used test
strips, empty the box.
T131
Photometric section
driving trouble
Содержание AUTION ELEVEN AE-4022
Страница 1: ......
Страница 75: ...Chapter 2 Measurement Operation AE 4022 2 40 ...
Страница 112: ...3 7 SETUP user settings AE 4022 3 37 4 Quit the SETUP menu aPress twice The display returns to the Standby screen ...
Страница 143: ...Chapter 4 Maintenance AE 4022 4 14 5 Close the printer cover aClose the printer cover Printer cover ...
Страница 146: ...AE 4022 5 1 Chapter 5 Troubleshooting 5 1 Warning messages 5 2 5 2 Error messages 5 3 5 3 Trouble messages 5 5 ...
Страница 153: ...Chapter 5 Troubleshooting AE 4022 5 8 ...
Страница 160: ......
Страница 161: ......