001-31249-EN rev
. 3
Care and Maintenance
38
Caring for Accessories
Other than cleaning, no user maintenance is required, but some components may need to be serviced or
replaced periodically. This should be carried out by a qualified Arjo service technician. For more
information contact Arjo or an approved distributor.
Handling, Transport and Storage
Avoid violent impacts while transporting the ceiling lifts.
The ceiling lifts should not remain stored for long periods of time without recharging the batteries.
NOTE: Even if the lift is not used, Arjo recommends charging the batteries at least every 2 weeks. This will
prevent premature aging of batteries.
If you store or ship the
Maxi Sky
2 Plus, ensure that the power is turned off (no green light) beforehand.
Battery Replacement
For more information regarding the ceiling lift batteries and their replacement information, refer to the
Maxi Sky 2 Instructions for Use
(001-15698-
XX
).
To replace batteries, be sure to contact your local Arjo agent.
CAUTION:
Do not attempt to use a battery that was not supplied by Arjo. Batteries are specially
designed for Arjo charging systems. Attempting to use an unauthorized battery may seriously damage
the lift and/or the charger.
Annual Inspection
The
Maxi Sky
2 Plus trolley, ceiling lifts and accessories must be inspected every 12 months as a
minimum requirement. Do not attempt to perform the inspection unless you are qualified to do so. Refer
to the "Preventive Maintenance Schedule" on page 39.
For information regarding the maintenance schedule for the ceiling lifts that are part of the
Maxi Sky
2 Plus, please refer to the
Maxi Sky 2 Instructions for Use
(001-15698-
XX
). Special attention
should be brought to the ceiling lift straps.
Maintenance Requirements
The ceiling lifts of the
Maxi Sky
2 Plus are equipped with a maintenance monitor which informs the user
that a maintenance inspection is necessary. The screen will display the "maintenance required" (
)
symbol to draw the user's attention.
Arranging for scheduled inspections ensures the durability of the unit and the security of the patient and
user.
For further information about the maintenance of your
Maxi Sky
2 Plus trolley and spreader bar contact
your Arjo agent.