001-31249-EN rev
. 3
Technical Specifications
46
Technical Data
product information
m
axi
s
ky
2 p
lus
Weight, complete (incl. trolley and two ceiling lifts)
51.6 kg (113.5 lb)
Weight, spreader bar
6.6 kg (14.5 lb)
Lifting capacity
Dual Mode 454 kg (1000 lb), Single Mode 272 kg (600 lb)
Strap length
2500 mm (98.4 in)
Lifting speed (Dual Mode) 4.3 cm/s (1.7 in/s) without load
3.4 cm/s (1.3 in/s) at 240 kg (528 lb)
3.0 cm/s (1.2 in/s) at 400 kg (880 lb)
2.9 cm/s (1.1 in/s) at 454 kg (1000 lb)
Maximum stroke (from ceiling)
2300 mm (90.6 in)
Horizontal displacement speeds
N/A, Manually operated
Operating force of control
< 5 N
Push and pull forces at 454 kg (1000 lb)
Horizontal Movement and Rotation
< 60 N
electrical
Duty cycle
Up/Down Motor: Max 10%, 1 minute continuously
Sit/Reclined Motor (Single Mode): Max 10%, 1 minute continuously
Horizontal displacement speeds N/A
Vertical axis motor 24 VDC, 110 watts
Rating 24 VDC, 15 A max
Noise level for either raising or lowering, with or without load 58.2 dBA with no load, 65.3 dBA at 454 kg (1000 lb)
Medical equipment Type BF protection against electrical shock in accordance with IEC 60601-1
WARNING:
Portable RF communications equipment (including peripherals such as antenna cables and external antennas)
should be used no closer than 30 cm to any part of the Maxi Sky 2 Plus, including cables specified by the manufacturer.
Otherwise, performance degradation of this equipment could result. See the "Electromagnetic Compliance" section for more
details.
Battery type 12 V, 5 Ah, sealed rechargeable valve regulated lead acid battery
Battery capacity
Provides up to 110 transfers with a load of 200 kg (440 lb), up to 50 transfers with a
load of 400 kg (880 lb) and up to 34 transfers with a load of 454 kg (1000 lb)
Degree of protection - hand control IPX7
Degree of protection - Maxi Sky 2 Plus IP24
Ceiling lift - protection class - shock prevention Internally powered equipment
Battery charger input 100-240 VAC, 50-60 Hz, 57 – 70 VA
Battery charger output 28.1 VDC, 1 A max, 28.1 VA
Battery Charger safety protection Class 2, double insulated
operation, transport and storage conditions
Normal Operation
Ambient temperature range: 5°C to 40°C (41°F to 104°F)
Relative humidity range: 15% to 93%
Atmospheric pressure range: 700 hPa to 1060 hPa
Transport / Storage
Ambient temperature range: -40°C to 70°C (-40°F to 158°F)
Relative humidity range: 10% to 93% non-condensing
Atmospheric pressure range: 500 hPa to 1060 hPa
WARNING: This equipment is not suitable in the presence of flammable anesthetic mixtures with air or oxygen, or with nitrous
oxide.