Care and Maintenance
37
Care and Maintenance
Disinfection
Wipe Maxi Sky 2 PLUS & accessories using a solution containing one of the compatible disinfectants below.
Disinfection methods should comply with local or national guidelines (Decontamination of Medical Devices),
depending on Healthcare Facility or country of use.
If uncertain, consult local Infection Control specialist.
NOTE: It is recommended to clean the lift and its accessories before disinfection.
Disinfection Agents for
Lift and Accessories
Disinfection Agents
Quaternary Ammonium ≤ 0.21% (2,100 ppm)
Isopropyl Alcohol ≤ 70%
Phenolic ≤ 1.56% (15,600 ppm)
Bleach ≤ 1% (10,000 ppm)
Peracetic Acid ≤ 0.25% (2,500 ppm)
Hydrogen Peroxide ≤ 1.5%
Standard model (woven nylon strap & Spreader Bar with padding)
Infection Control model (coated strap & Spreader Bar without padding)
*
Other Accessories for all models
**
* Some parts on the spreader bar may discolour after repeated contact with bleach, peracetic acide or hydrogen peroxide.
** Handset attachment cable may become slightly grimy after repeated contact with bleach or hydrogen peroxide.
NOTE: After thorough cleaning and/or disinfection of the Emergency Cord, assure the lift is still "ON".