6RL SERIES OPERATION MANUAL
SECTION 6: FRONT PANEL OPERATION
Entire Contents Copyright
2018 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS.
6RL Series Regenerative DC Load Operation Manual
Page 93 of 204
2.
In the menu tap on the touch area
and then on the desired function.
Note: this touch area is locked in R mode (adjustable resistance).
3.
Depending on the choice of function there follows a request to which value the
function generator is going to be applied:
or
.
4.
Adjust the parameters as you desire, like offset, amplitude and frequency for a sine
wave, for example.
5.
Adjust the overall limits of voltage, current and power, which you can access with
touch area
.
Note: When entering function generator mode, these limits are reset to safe values, which
can prevent the function from working at all. For example, if you apply the selected
function to the output current, then the overall current limit should not interfere
and should at least be as high as amplitude.
How to Load a Function:
1.
After setting the values for the required signal
generation, tap on the touch area
.
The unit will then load the data into the internal controller
and changes the display. Shortly afterwards the static
values are set (power and voltage or current), the DC Input
is switched on and the touch area
is enabled. Only then can the function output
be started.
Note: The static values are applied to the DC Input immediately after loading the function
as it switches the DC Input on automatically in order to set the start condition. They
represent start and end values for the progress of the function, so that it does not
need to start from 0. Only exception: when applying a function to the current (I),
there is no adjustable static current value, so the function will always start from 0 A
How to Start and Stop a Function:
1.
The function can be started either by tapping
or pushing the “On/Off” button, if the
DC Input is currently switched off. The function
then starts immediately. In case START is used
while the DC Input is still switched off, it will be
switched on automatically.
2.
The function can be stopped either by tapping
or operating the
“On/Off” button. However, there is a difference: