6RL SERIES OPERATION MANUAL
SECTION 7: REMOTE PROGRAMMING
Entire Contents Copyright
2018 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS.
6RL Series Regenerative DC Load Operation Manual
Page 148 of 204
7.12.6.3
[SOURce:]POWer
˽
<NRf>[Unit]
Sets the input resp. output power limit of the unit within a certain range, which is either
defined by adjustment limits ("Limits", where featured) or is 0...102% nominal value, or
reads the last setting. Alternatively, parameters MIN or MAX can be used to instantly set the
power to zero (MINimum) or MAXimum.
Query form:
[SOURce:]POWer?
Value range:
<NRf> = 0...1.02 * nominal power (according to technical specs)
Examples:
POW˽3000
Absolute short form. Sets 3000 W, unless the power
adjustment range has been limited otherwise.
SOUR:POWER˽3.5kW
Mixed form short/long, with unit and magnitude Kilo. Sets
3.5 kW resp. 3500 W, unless the adjustment range of the
power has been limited otherwise.
SOURCE:POWER˽MIN
Sets the power to the defined minimum, which is usually 0
W.
7.12.6.4
[SOURce:]RESistance
˽
<NRf>[Unit]
Sets the output power limit of the unit within a certain range, which is either defined by
adjustment limits ("Limits", where featured) or is 0...102% nominal value, or reads the last
setting. Alternatively, parameters MIN or MAX can be used to instantly set the power to
zero (MINimum) or MAXimum.
Query form:
[SOURce:]POWer?
Value range:
<NRf> = 0...1.02 * nominal power (according to technical specs)
Examples:
POW˽3000
Absolute short form. Sets 3000 W, unless the power
adjustment range has been limited otherwise.
SOUR:POWER˽3.5kW
Mixed form short/long, with unit and magnitude Kilo. Sets
3.5 kW resp. 3500 W, unless the adjustment range of the
power has been limited otherwise.
SOURCE:POWER˽MIN
Sets the power to the defined minimum, which is usually 0
W.
7.12.6.5
[SOURce:]RESistance
˽
<NRf>[Unit]
Power supplies with internal resistance feature use this value to simulate an internal resistor
in series to the output, where the output voltage differs from the adjusted value by an
amount that calculates from the adjusted resistance value and actual output current. The
way of setting the resistance value on both unit types is identical. The adjustable range can
be limited with an upper adjustment limit. Alternatively, parameters MIN or MAX can be
used to instantly set the resistance to the adjustable MINimum or MAXimum.
Query form:
[SOURce:]RESistance?