TECHNICAL SPECIFICATIONS
192 / 194 NTA47224-300A9
CALYS 150
Frequency and pulses
Range
Resolution
Range
Accuracy / 1 year
Notes
1 000 Hz
0.01 Hz
0.01 Hz to 1 000 KHz
0.005% R + 5 mHz
100 kHz
1 Hz
1 Hz to 100 KHz
0.005% R + 5 mHz
Temperature coefficient < 5 ppm L/°C from 0°C to +18°C and from +28°C to +50°C.
Additional characteristics in simulation
D.2.7.1
Generation of increments
The increment generation function is used to program an incremental progression of the active transmission function.
D.2.7.2
Generation of ramps
The ramp generation function is used to program a linear variation of the active transmission function.
D.2.7.3
Synthesizer
The synthesiser function is used:
to store up to 100 transmission values in permanent memory,
to recall and transmit manually or automatically the contents of these memories.
D.2.7.4
Scale correction
The scale correction function performs a conversion between the physical quantities displayed and the electrical quantities
simulated.