GETTING STARTED
146 / 194 NTA47224-300A9
CALYS 150
When a sensor supports this function, the screen is linked to the sensor.
B.2.11.6 Device status
The "Device status" item from the HART menu is used to access the following: Device status, Clear modified configuration
indication and Reset device.
To access Device status, select "Device status", "Device status" items from the HART menu.
A table displays some device status registers:
-
Device malfunction: indicates device malfunction. Refer to device user manual and check for compatibility with
current operation.
-
Modified configuration: this check box indicates that configuration has been modified. This information may be reset
in the next chapter.
-
Cold start: indicates a cold start???
-
More detailed status available: indicates that a more detailed status is available???
-
Fixed current loop: indicates that curred is fixed, which is not the case in point-to-point-mode for instance.
-
Saturated current loop: check connections in accordance with equipment manual.
-
Variable other than PV out of range:
-
Primary variable (PV) out of range:
Once status assessed, press the F1 key ("OK") to switch back to normal operation.