GETTING STARTED
122 / 194 NTA47224-300A9
CALYS 150
In measurement mode, channel V1 is connected to the four "IN" terminals located on the left half of the unit:
Measuring DC voltage
Display the MEASUREMENT MENU dialogue box:
Select the Vdc measurement function then the correct measurement range using the function and navigation keys.
Confirm with ENTER.
The following ranges are available:
Range
100 mV
1 V
10 V
50 V
Auto
Resolution
1 µV
10 µV
100 µV
1 mV
---
Input impedance
> 1 G
Ω
> 1 G
Ω
10 M
Ω
10 M
Ω
(1)
---
(1) WEM41012-00B
10 M
Ω
; WEM41012-00A
1 M
Ω
;
The DC voltage to be measured is connected between the V and COM terminals.