ADVANCED OPERATION
160 / 194 NTA47224-300A9
CALYS 150
Staircase mode
This mode is used to program an incremental progression of the active transmission function.
When this mode is active, the
icon appears in the transmission window.
The
function key launches a cycle of increasing increments and the
function key launches a cycle of decreasing
increments.
The default parameters of this mode are displayed on the left hand side of the transmission window. To change these
parameters, press ENTER or use the Menu
Mode.
The parameters of a staircase are:
B: minimum amplitude of the signal.
H: maximum amplitude of the signal.
I: amplitude of the increment
T: duration of the steps in seconds
Delay: delay in seconds between launching the staircase and transmission of the first increment.
During generation of a staircase, a progress bar indicates the state of progress. The function keys control generation:
The
key stops generation at any time
The
key suspends generation
The
key commences or resumes generation
An
icon in the transmission window indicates suspended generation.