"Transmission / Simulation" Function
D
NTA47224-300A9
189 / 194
D.2 "Transmission / Simulation" Function
Maximum rated voltage in common mode: 60 VDC or VAC.
DC Voltage
Range
Resolution
Measuring range
Accuracy / 1 year
Notes
+100 mV
1 µV
- 5 mV to + 100 mV
0.005% R + 2 µV
Output load min = 1 KOhm
+1 V
10 µV
- 5 mV to + 1 V
0.005% R + 8 µV
Output load min = 2 KOhm
+10 V
100 µV
- 100 mV to + 10 V
0.007% R + 80 µV
Output load min = 4 KOhm
+50 V
1 mV
- 100 mV to + 50 V
0.007% R + 0.5 mV
Output load min = 4 KOhm
Temperature coefficient < (5 ppm L + 0.5 ppm Cal)/°C from 0°C to +18°C and from +28°C to +50°C.
DC Current
Range
Resolution
Accuracy / 1 year
Notes
24 mA
1 µA
0.007% R + 0.8 µA
4-20 mA
1 µA
0.007% R + 0.8 µA
0-20 mA
1 µA
0.007% R + 0.8 µA
Temperature coefficient < (6 ppm L + 0.5 ppm Cal)/°C from 0°C to +18°C and from +28°C to +50°C.
The above specifications are given for the following configurations:
- CALYS 150 active mode (+24 V ON)
Passive mode measurer (+24 V OFF).
- CALYS 150 passive mode (+24 V OFF)
Active mode measurer (+24 V ON).
Resistance
Range
Resolution
Measuring range
Accuracy / 1 year
Notes
400
Ω
- 1 mA
(direct current)
10 m
Ω
1
Ω
to 400
Ω
0.006% R + 20 m
Ω
Iext of 0.1 mA / 1 mA
400
Ω
- 1 mA
(Pulsed current)
10 m
Ω
1
Ω
to 400
Ω
0.006% R + 30 m
Ω
Iext of 0.1 mA / 1 mA
400
Ω
- 4 mA
(direct current)
10 m
Ω
1
Ω
to 400
Ω
0.006% R + 20 m
Ω
Iext of 1 mA / 4 mA
400
Ω
- 4 mA
(Pulsed current)
10 m
Ω
1
Ω
to 400
Ω
0.006% R + 30 m
Ω
Iext of 1 mA / 4 mA
3600
Ω
(direct current)
100 m
Ω
10
Ω
to 3600
Ω
0.006% R + 100 m
Ω
Iext of 0.1 mA / 1 mA
3600
Ω
(Pulsed current)
100 m
Ω
10
Ω
to 3600
Ω
0.006% R + 200 m
Ω
Iext of 0.1 mA / 1 mA
Accuracy applies to a 4-wire connection with Y-shaped lugs, add 20 mOhm of uncertainty when banana plugs are used.
Temperature coefficient < (5 ppm L + 0.5 ppm Cal)/°C from 0°C to +18°C and from +28°C to +50°C.