GETTING STARTED
142 / 194 NTA47224-300A9
CALYS 150
The CALYS 150 then runs the various steps defined for verification step-by-step. The procedure may be accelerated by
pressing the F3 key ("Stable") when stability is deemed satisfactory.
At the end of the verification procedure, the results may be "Saved" with the F1 key or exited with the "ENT" key.
The unit displays the following prompt: "Save calibration report before closing?". Press keys F1 ("YES") or F2 ("NO").