GETTING STARTED
154 / 194 NTA47224-300A9
CALYS 150
B.3.1.7
Frequency generation
Display the TRANSMISSION CONFIGURATION dialogue box:
Select the Frequency transmission function, then the range using the function and navigation keys.
Select the Signal “Output type”.
Enter cyclic ratio between 20 and 80%
Enter the amplitude of the signal between 0 and 20 V.
Confirm with ENTER.
The following ranges are available:
Range
1,000 Hz
10,000 Hz
Resolution
0.01 Hz
0.1 Hz
Max. Amplitude
20 V
20 V
Depending on the range selected, display is in Hz or beats per minute (BPM).
Connection of the frequency source generated is between the Hz and COM terminals.
B.3.1.8
Frequency generation for dry contact
Display the TRANSMISSION CONFIGURATION dialogue box:
Select the Frequency transmission function, then the range using the function and navigation keys.
Select the “Type of Output” Hard contact.
Enter cyclic ratio between 20 and 80%
Confirm with ENTER.
The following ranges are available:
Range
1,000 Hz
10,000 Hz
Resolution
0.01 Hz
0.1 Hz
Max. Amplitude
20 V
20 V
Depending on the range selected, display is in Hz or beats per minute (BPM).
Connection of the frequency source generated is between the Hz and COM terminals.