GETTING STARTED
144 / 194 NTA47224-300A9
CALYS 150
To create a verification procedure for the detector part, press the F1 key ("YES") or else cancel by pressing the F2 key
("NO").
When a procedure is created, the letter "H" in the HART column indicates that it is a verify detector procedure for the detector
and current loop. Edit procedure to complete creation.
From the "HART" item, "Verify", "Define procedure", the three types of verification procedure may be directly edited: Loop,
Detector, De Loop. See relevant sections in the above explanations.
B.2.11.5 Adjustment
The current loop and/or detector may be adjusted.
To adjust the current loop, select "Adjust", "Current loop" from the "HART" menu.