Calibrated sensors
C
NTA47224-300A9
169 / 194
To enter calibration points in the table, use the
button.
Enter the values and press ENTER.
Use the following buttons to continue configuring a sensor:
to edit an existing calibration point,
to add a calibration point,
to delete a calibration point.
Between 1 and 4 calibration points can be entered per sensor.
These calibration points are used to calculate a polynomial c(T) of degree 0 to 3, giving the sensor's voltage (or resistance)
correction at temperature T.
In the specific case where a single calibration point is given, the behavior will differ according to whether the sensor is a
thermocouple or a thermometric resistance:
In the case of a thermocouple, the correction is a fixed voltage deviation.
In the case of a resistive probe, the correction made is an R0 correction.
Sensors declared in this manner are added to the list of couple types (or of probe types) proposed in the measurement
function settings dialog box. They appear at the top of the list, in front of the standard sensors. Their name is preceded by
the '*' character, indicating that these are calibrated sensors.