Generation / Simulation or Measurement on Channel V2
B
NTA47224-300A9
155 / 194
B.3.1.9
Pulse generation
Display the TRANSMISSION CONFIGURATION dialogue box:
Select the Pulse transmission function, then the range using the function and navigation keys.
Select the “Type of Output” Signal.
Enter the amplitude of the signal between 0 and 20 V.
Confirm with ENTER.
The
icon appears in the Transmission window.
To change the default parameters, press ENTER or enter the Configuration … menu then Pulses ….
The parameters are:
Number: Number of pulses to be generated between 1 and 8,000,000.
Overall time: burst time in milliseconds (<10Hz) between 0.05 and 100 000 000 or in seconds (>10Hz) between
0.02 and 1 000 000
Cyclic ratio: cyclic ratio between 20 and 80 %
To commence pulse generation, press Start (F3).