Simulation Modes
C
NTA47224-300A9
161 / 194
Simple Ramp Mode
The simple ramp generation function is used to program a linear variation in one direction (increasing or decreasing) of the
active transmission function.
When this mode is active, the
icon appears ion the transmission window.
The
key is used to launch an increasing ramp and the
function key is used to launch a decreasing ramp.
The default parameters of this mode are displayed in the left hand side of the transmission window. To change these
parameters, press ENTER or use the Menu
Mode….
The parameters of a simple ramp are:
B: minimum amplitude of the signal.
H: maximum amplitude of the signal.
T: duration of the ramp in seconds.
Delay: delay in seconds between launching the ramp and the start of transmission.
During generation of a simple ramp, a progress bar indicates the state of progress. The function keys are used to control
generation:
The
key stops generation at any time
The
key suspends generation
The
key commences or resumes generation
An
icon in the transmission window indicates suspended generation.