Measurement Function (Channel 1 and Channel 2)
D
NTA47224-300A9
187 / 194
The precision is guaranteed for a reference junction temperature of 0°C.
When using the internal reference junction (except couple B) add an additional uncertainty of 0.2°C at 0°C. For other
temperatures, account must be taken of the sensitivity of the thermocouple to the temperature (T) in question, giving an
additional uncertainty of 0.2°C * S(0°C)/S(T).
Temperature coefficient: < 10 % of the accuracy/°C from 0°C to +18°C and from +28°C to +50°C.
Display in °C, °F and K.
It is possible, thermocouple B excepted, to choose the location of the cold junction by programming from the keyboard:
external at 0°C,
internal (compensation for the temperature of the terminals of the unit).
by programming the temperature.
Temperature using resistive probes
Type of sensors:
Pt 50ohm, 100ohm, 200ohm, 500ohm, 1,000ohm with
α
= 3851 as per CEI 751/1995
Pt 100ohm with
α
= 3916 as per JIS C 1604/1989
Pt 100ohm with
α
= 3926 as per EIT90
Ni 100ohm, 1,000 ohm with
α
=618 as per DIN 43760
Ni 120ohm with
α
=672 as per MIL-T-24388
Cu 10ohm with
α
=427 as per MINCO 16/9
Cu 50ohm with
α
=428 as per OIML R 84
Sensor
Measuring range
Resolution
Accuracy / 1 year
Pt 50 (
α
= 3851)
- 220°C to + 850°C
0,01°C
0.006% R + 0.04°C
Pt 100 (
α
= 3851)
- 220°C to + 850°C
0,01°C
0.006% R + 0.03°C
Pt 100 (
α
= 3916)
- 200°C to + 510°C
0,01°C
0.006% R + 0.03°C
Pt 100 (
α
= 3926)
- 210°C to + 850°C
0,01°C
0.006% R + 0.03°C
Pt 200 (
α
= 3851)
- 220°C to + 850°C
0,01°C
0.006% R + 0.04°C
Pt 500 (
α
= 3851)
- 220°C to + 850°C
0,01°C
0.006% R + 0.03°C
Pt 1000 (
α
= 3851)
- 220°C to + 740°C
0,01°C
0.006% R + 0.03°C
Ni 100 (
α
= 618)
- 60°C to + 180°C
0,01°C
0.006% R + 0.05°C
Ni 120 (
α
= 672)
- 40°C to + 205°C
0,01°C
0.006% R + 0.05°C
Ni 1000 (
α
= 618)
- 60°C to + 180°C
0,01°C
0.006% R + 0.05°C
Cu 10 (
α
= 427)
- 50°C to + 150°C
0,10°C
0.006% R + 0.18°C
Cu 50 (
α
= 428)
- 180°C to + 200°C
0,01°C
0.006% R + 0.05°C
For negative temperatures, use the displayed value L and not its absolute value.
Temperature coefficient: < 10 % of the accuracy/°C from 0°C to +18°C and from +28°C to +50°C.
The above accuracy is given for 4 wire connection of the temperature sensor ("Y shaped lug" connection).
The above accuracy takes into account the intrinsic error of the temperature sensor used and its conditions of use.