GETTING STARTED
152 / 194 NTA47224-300A9
CALYS 150
B.3.1.4
Resistive probe simulation (temperature)
Display the TRANSMISSION CONFIGURATION dialogue box:
Select the Rt transmission function, then the appropriate “type of probe”, and range using the function and
navigation keys.
Select the display unit
Select measurement current (direct or pulsed)
Confirm with ENTER.
Connection is made between the two
Ω
terminals.
The following probes are available:
Sensor
Probe Type Caption
Pt 50 (
α
= 3851)
Pt 50
Pt 100 (
α
= 3851)
Pt 100
Pt 100 (
α
= 3916)
Pt 100-3916
Pt 100 (
α
= 3926)
Pt 100-3926
Pt 200 (
α
= 3851)
Pt 200
Pt 500 (
α
= 3851)
Pt 500
Pt 1000 (
α
= 3851)
Pt 1000
Ni 100 (
α
= 618)
Ni 100
Ni 120 (
α
= 672)
Ni 120
Ni 1000 (
α
= 618)
Ni 1000
Cu 10 (
α
= 427)
Cu 10
Cu 50 (
α
= 428)
Cu 50
The resistive temperature probe simulation function can be used with either 2, 3 or 4 wire connection.
As for the resistance simulation function, if a polling acquisition system is used, ensure that the transmitter carries
out the measurement at least 1 ms after the current is present.