5-14 | IPS-4 Dual Bench (UV / IR) Analyzer
5.
GP Analyzers Only:
Turn off the Instrument Air supply to the analyzer and then close the
Aspirator Air valve.
Purged Analyzers:
The Instrument Air supply must always remain on for the Purge By-
pass Switch to operate properly.
6. Removing/cleaning parts in the Measuring Cell:
The analyzer sample system will be under positive pressure.
Various lengths of screws are encountered while disassembling the
Measuring Cell Assembly. Take note of where each length is used and
be sure to replace the screws in their original locations when reassem-
bling these parts later.
a. Disconnect the Sample Line and Vent Line tubes from the Cell.
Loosen and remove the Cell RTD (if used) from the RTD Clamp
(not shown) on the Cell.
Disconnect all other sample system parts/tubing to allow removal
of the Cell.
b. Note the position of the Cell and its Sample/Vent fittings and Re-
flector Block.
Remove the three screws that secure the Cell to the Window
Housing Assembly (not shown). Carefully pull the Cell straight out
from the Window Housing and place it on a work bench or other
suitable location to work on it.
If the Oven/Sample System Enclosure is heated
, close its door to
help maintain the temperature inside the enclosure while working
on the Cell.
In the next steps use a soft, non-abrasive cloth to place the Cell parts
on.
c. If the Measuring Cell has an Optical Bench/Cell Insulator, remove
the three screws that secure it to the Cell and remove the Insulator.
NOTE
NOTE
!
WARNING
Содержание Western Research IPS-4
Страница 48: ...2 18 IPS 4 Dual Bench UV IR Analyzer This page intentionally left blank...
Страница 140: ...4 48 IPS 4 Dual Bench UV IR Analyzer This page intentionally left blank...
Страница 172: ...5 32 IPS 4 Dual Bench UV IR Analyzer This page intentionally left blank...
Страница 178: ...6 6 IPS 4 Dual Bench UV IR Analyzer This page intentionally left blank...
Страница 218: ...A 2 IPS 4 Dual Bench UV IR Analyzer Analyzer Light Path UV Optical Bench Schematic...
Страница 219: ...Appendix A Drawings A 3 Analyzer Light Path NDIR Optical Bench Schematic CONDENSING...
Страница 222: ...A 6 IPS 4 Dual Bench UV IR Analyzer Detector Board 100 2046...
Страница 223: ...Appendix A Drawings A 7 Display Interface Board 100 2049...
Страница 225: ...Appendix A Drawings A 9 AC Disconnection Board 100 2077...
Страница 227: ...Appendix A Drawings A 11 Relay Board 100 2050...
Страница 228: ...A 12 IPS 4 Dual Bench UV IR Analyzer Analog Board 100 2047...
Страница 229: ...Appendix A Drawings A 13 MCU Board 100 2045...
Страница 230: ...A 14 IPS 4 Dual Bench UV IR Analyzer Xenon Lamp Power Supply Board 100 2061 UV Optical Bench...
Страница 231: ...Appendix A Drawings A 15 Infrared Source Assembly Interface Board 100 2838 IR Optical Bench...
Страница 232: ...A 16 IPS 4 Dual Bench UV IR Analyzer Electronics Enclosure Wiring Diagram Sheet 1 of 5...
Страница 233: ...Appendix A Drawings A 17 Electronics Enclosure Wiring Diagram Sheet 2 of 5...
Страница 234: ...A 18 IPS 4 Dual Bench UV IR Analyzer Electronics Enclosure Wiring Diagram Sheet 3 of 5...
Страница 235: ...Appendix A Drawings A 19 Disconnect Enclosure Wiring Diagram Sheet 4 of 5...
Страница 236: ...A 20 IPS 4 Dual Bench UV IR Analyzer Disconnect Enclosure Wiring Diagram Sheet 5 of 5...
Страница 238: ...S 2 IPS 4 Dual Bench UV IR Analyzer This page intentionally left blank...