Controller / User Interface
| 4-21
Calibration Actions Screen
The
Calibration action
screen displays all of the available analyzer calibra-
tion functions and allows the user to initialize the calibration functions.
Figures 4-17a and 4-17b illustrate the differences between analyzers that
use single-species spanning and those that use multi-species spanning.
Start zero calibration
This function forces the analyzer into the Zero calibration portion of
the normal operating sequence, starting with the “
Zero Flush
” mode,
then “
Zero
.” The analyzer zero is automatically adjusted based on the
average reading during the zero calibration.
Select Span Fluid (for applications that require multiple span fluids)
This function is used in applications where multiple span fluids are
required. The span fluid connected to the
Cal (Span) Fluid Inlet
port
must correspond to the fluid selected from the options listed under
this parameter.
To select a span fluid, press
ENTER
to go into edit mode. Use the
UP
/
DOWN
arrow key to select the fluid you wish to span and then press
ENTER
to select the span fluid. Next, select
Start span calibration
and
press
ENTER
to start the span.
Start span calibration
This function forces the analyzer into the Span mode of the normal
operating sequence, starting with the “
Span Fluid Flush
” mode. The
analyzer span (calibration) is automatically adjusted based on an inte-
grated reading during the span calibration. After the span is complete,
the analyzer will return to normal operation (
Measure
mode).
Figure 4-17a.
Calibration screen,
applications with
single span fluid.
►
Start Zero Calibration
Start Span Calibration
Start Zero CaliSpan Calibration
Inject Zero Fluid
Inject Span Fluid
Start Measure
HOME ALARMS STATUS SETUP
Figure 4-17b.
Calibration screen,
applications with
multi-span fluids.
►
Start Zero calibration
Select Span Fluid
Start Apan Calibration
Start Zero+Span Calibration
Inject Zero Fluid
Inject Span Fluid
HOME ALARMS STATUS SETUP
Содержание Western Research IPS-4
Страница 48: ...2 18 IPS 4 Dual Bench UV IR Analyzer This page intentionally left blank...
Страница 140: ...4 48 IPS 4 Dual Bench UV IR Analyzer This page intentionally left blank...
Страница 172: ...5 32 IPS 4 Dual Bench UV IR Analyzer This page intentionally left blank...
Страница 178: ...6 6 IPS 4 Dual Bench UV IR Analyzer This page intentionally left blank...
Страница 218: ...A 2 IPS 4 Dual Bench UV IR Analyzer Analyzer Light Path UV Optical Bench Schematic...
Страница 219: ...Appendix A Drawings A 3 Analyzer Light Path NDIR Optical Bench Schematic CONDENSING...
Страница 222: ...A 6 IPS 4 Dual Bench UV IR Analyzer Detector Board 100 2046...
Страница 223: ...Appendix A Drawings A 7 Display Interface Board 100 2049...
Страница 225: ...Appendix A Drawings A 9 AC Disconnection Board 100 2077...
Страница 227: ...Appendix A Drawings A 11 Relay Board 100 2050...
Страница 228: ...A 12 IPS 4 Dual Bench UV IR Analyzer Analog Board 100 2047...
Страница 229: ...Appendix A Drawings A 13 MCU Board 100 2045...
Страница 230: ...A 14 IPS 4 Dual Bench UV IR Analyzer Xenon Lamp Power Supply Board 100 2061 UV Optical Bench...
Страница 231: ...Appendix A Drawings A 15 Infrared Source Assembly Interface Board 100 2838 IR Optical Bench...
Страница 232: ...A 16 IPS 4 Dual Bench UV IR Analyzer Electronics Enclosure Wiring Diagram Sheet 1 of 5...
Страница 233: ...Appendix A Drawings A 17 Electronics Enclosure Wiring Diagram Sheet 2 of 5...
Страница 234: ...A 18 IPS 4 Dual Bench UV IR Analyzer Electronics Enclosure Wiring Diagram Sheet 3 of 5...
Страница 235: ...Appendix A Drawings A 19 Disconnect Enclosure Wiring Diagram Sheet 4 of 5...
Страница 236: ...A 20 IPS 4 Dual Bench UV IR Analyzer Disconnect Enclosure Wiring Diagram Sheet 5 of 5...
Страница 238: ...S 2 IPS 4 Dual Bench UV IR Analyzer This page intentionally left blank...