Installation and Start-Up
| 3-17
Installing and Connecting the Purge Fluid Line
Instrument Air or Nitrogen can be used to operate the purge system. This
procedure assumes that Instrument Air is being used and is the same for
Nitrogen. If using instrument air, the air must meet the ANSI/ISA-7.0.01
(1996) specification. If using nitrogen, use only ultrahigh purity (UHP)
nitrogen.
The purge gas must be supplied via a pressure regulator and maintained
at 60 to 115 psig (4 to 8 barg) and must be able to provide a flow of at least
350 L/min (after the initial purge, the flow is very low – just enough to
compensate for leaks from the enclosure).
1. Provide a source of either Instrument Air or UHP grade Nitrogen.
2. Before connecting the purge inlet, purge the air line to clear any water
or particulates.
3. Connect the air supply to the 1/2" purge supply connection fitting on
the top of the analyzer (see Figure 3-4).
4. Ensure that the Purge Bypass Switch is in the “
ACTIVE
” position and
that all enclosure doors are closed and latched.
5. Open the purge inlet valve and set the regulator to 60 to 65 PSI
(4 barg). The left indicator should change from RED to GREEN and
the right from BLACK to YELLOW.
6. After approximately 5 minutes, the right indicator should change from
YELLOW to BLACK indicating a successful purge.
7. Open the Electronics Enclosure. The indicators on the purge controller
should go back to RED and YELLOW.
8. Close the inlet valve and open the door and proceed with the analyzer
installation.
The purge system should only be placed into BYPASS if the area is
known to be clear of flammable gases.
!
WARNING
Содержание Western Research IPS-4
Страница 48: ...2 18 IPS 4 Dual Bench UV IR Analyzer This page intentionally left blank...
Страница 140: ...4 48 IPS 4 Dual Bench UV IR Analyzer This page intentionally left blank...
Страница 172: ...5 32 IPS 4 Dual Bench UV IR Analyzer This page intentionally left blank...
Страница 178: ...6 6 IPS 4 Dual Bench UV IR Analyzer This page intentionally left blank...
Страница 218: ...A 2 IPS 4 Dual Bench UV IR Analyzer Analyzer Light Path UV Optical Bench Schematic...
Страница 219: ...Appendix A Drawings A 3 Analyzer Light Path NDIR Optical Bench Schematic CONDENSING...
Страница 222: ...A 6 IPS 4 Dual Bench UV IR Analyzer Detector Board 100 2046...
Страница 223: ...Appendix A Drawings A 7 Display Interface Board 100 2049...
Страница 225: ...Appendix A Drawings A 9 AC Disconnection Board 100 2077...
Страница 227: ...Appendix A Drawings A 11 Relay Board 100 2050...
Страница 228: ...A 12 IPS 4 Dual Bench UV IR Analyzer Analog Board 100 2047...
Страница 229: ...Appendix A Drawings A 13 MCU Board 100 2045...
Страница 230: ...A 14 IPS 4 Dual Bench UV IR Analyzer Xenon Lamp Power Supply Board 100 2061 UV Optical Bench...
Страница 231: ...Appendix A Drawings A 15 Infrared Source Assembly Interface Board 100 2838 IR Optical Bench...
Страница 232: ...A 16 IPS 4 Dual Bench UV IR Analyzer Electronics Enclosure Wiring Diagram Sheet 1 of 5...
Страница 233: ...Appendix A Drawings A 17 Electronics Enclosure Wiring Diagram Sheet 2 of 5...
Страница 234: ...A 18 IPS 4 Dual Bench UV IR Analyzer Electronics Enclosure Wiring Diagram Sheet 3 of 5...
Страница 235: ...Appendix A Drawings A 19 Disconnect Enclosure Wiring Diagram Sheet 4 of 5...
Страница 236: ...A 20 IPS 4 Dual Bench UV IR Analyzer Disconnect Enclosure Wiring Diagram Sheet 5 of 5...
Страница 238: ...S 2 IPS 4 Dual Bench UV IR Analyzer This page intentionally left blank...