5-28 | IPS-4 Dual Bench (UV / IR) Analyzer
Condition/Alarm Name
Description and Suggested Corrective Action
Corrective Action (continued):
•
Take appropriate safety precautions, open the Electronics Enclosure
door, and:
- Check the fuse(s) for the zone(s) which caused the alarm (i.e.,
Bench Heater
or
Oven Heater
, if used). Refer to the Fuse Legend
at the back of the Electronics Enclosure for the locations of these
fuses.
•
Other checks/corrective action:
- Using an Ohm Meter, measure the
resistance of the associated
RTD (Cell, Bench, or Oven) and test it for an open circuit.
If the
RTD is faulty, contact AMETEK for assistance.
- Check the electrical connections between the heater and the
Relay
board. Check for proper connections and damage to the wiring.
- Replace the
Relay
board.
Contact AMETEK for assistance.
- Replace the
Customer Connection
board.
Contact AMETEK for
assistance.
For alarms caused by
High
temperatures only:
•
If the Oven Heater temperature-controlled zone has exceeded its
over-temperature limit of 177 °C/350 °F, the analyzer automatically
switches to
Zero Gas Flush
mode.
Take appropriate safety precautions, open the Electronics Enclosure
door, and:
- Manually reset the alarm by pressing the OT (Over-Temperature)
Reset Switch,
S501
, on the
Relay
board (see Figure 5-4) to re-
energize the tripped Oven Heater circuit.
- Check the LEDs for the Oven Heater Over-Temp circuit (
D502
on
the
Relay
board) to check if it is experiencing problems.
- Check the fuse for the Over-Temperature circuit (
F3
, located at
TB1
13
). Refer to the Fuse Legend for the locations of these fuses.
Pressure transducer 1
Pressure transducer 2
These alarm conditions indicate that the concentration of the output
exceeds its full-scale range by more than 10 psia (typical; set point and
pressure unit may vary).
Corrective Action:
•
The parameters have exceeded the normal operating range. From the
User Interface Panel, check that the full-scale range is correct for the
current sample. If
Pressure Transducer
alarms persist during normal
operation, contact the factory. A range change and recalibration may
be required.
Содержание Western Research IPS-4
Страница 48: ...2 18 IPS 4 Dual Bench UV IR Analyzer This page intentionally left blank...
Страница 140: ...4 48 IPS 4 Dual Bench UV IR Analyzer This page intentionally left blank...
Страница 172: ...5 32 IPS 4 Dual Bench UV IR Analyzer This page intentionally left blank...
Страница 178: ...6 6 IPS 4 Dual Bench UV IR Analyzer This page intentionally left blank...
Страница 218: ...A 2 IPS 4 Dual Bench UV IR Analyzer Analyzer Light Path UV Optical Bench Schematic...
Страница 219: ...Appendix A Drawings A 3 Analyzer Light Path NDIR Optical Bench Schematic CONDENSING...
Страница 222: ...A 6 IPS 4 Dual Bench UV IR Analyzer Detector Board 100 2046...
Страница 223: ...Appendix A Drawings A 7 Display Interface Board 100 2049...
Страница 225: ...Appendix A Drawings A 9 AC Disconnection Board 100 2077...
Страница 227: ...Appendix A Drawings A 11 Relay Board 100 2050...
Страница 228: ...A 12 IPS 4 Dual Bench UV IR Analyzer Analog Board 100 2047...
Страница 229: ...Appendix A Drawings A 13 MCU Board 100 2045...
Страница 230: ...A 14 IPS 4 Dual Bench UV IR Analyzer Xenon Lamp Power Supply Board 100 2061 UV Optical Bench...
Страница 231: ...Appendix A Drawings A 15 Infrared Source Assembly Interface Board 100 2838 IR Optical Bench...
Страница 232: ...A 16 IPS 4 Dual Bench UV IR Analyzer Electronics Enclosure Wiring Diagram Sheet 1 of 5...
Страница 233: ...Appendix A Drawings A 17 Electronics Enclosure Wiring Diagram Sheet 2 of 5...
Страница 234: ...A 18 IPS 4 Dual Bench UV IR Analyzer Electronics Enclosure Wiring Diagram Sheet 3 of 5...
Страница 235: ...Appendix A Drawings A 19 Disconnect Enclosure Wiring Diagram Sheet 4 of 5...
Страница 236: ...A 20 IPS 4 Dual Bench UV IR Analyzer Disconnect Enclosure Wiring Diagram Sheet 5 of 5...
Страница 238: ...S 2 IPS 4 Dual Bench UV IR Analyzer This page intentionally left blank...