3-30 | IPS-4 Dual Bench (UV / IR) Analyzer
7. While working from the User Interface Panel, perform the following
checks to ensure the analyzer temperature-control zone circuitry is
operating properly and the system pressures are set appropriately.
Record the temperature value for each zone and the system pressures
to use as a reference later while comparing the values.
Instructions for working from the “User Interface Panel”, refer to
Chapter 4 page 4-5.
If changes need to be made to any of the parameters settings, a pass-
word must be entered to allow the changes to be made. For details
about entering a password, refer to “Login/Set Passwords Screen” in
Chapter 4 page 4-30.
Figure 3-6.
STATUS menu
(example screen).
>
Heater
19 °C
Bench
24 °C
Cell
23 °C
Electronics
24 °C
Manifold Pressure
85 psia
Aspirator Pressure
3 psia
HOME
ALARMS
STATUS
SETUP
a. Press the
STATUS
soft key to view the
STATUS
screen (Figure 3-6).
Check and record the current operating temperature readings.
This will help you determine if the signal wiring is properly termi-
nated and if all of the circuitry is operating properly.
Pay close attention to the temperatures of the (Oven)
Heater
(if
used), (Optical)
Bench
, and
Cell
(if used) readings. If these tem-
peratures do not increase, check the wiring for proper connec-
tions. If the wiring is properly connected, it may indicate a shorted
RTD for that temperature zone.
The Oven Heater temperature parameter is available only for applica-
tions with a heated Oven.
b. Check the pressure of the pilot valves – Manifold block and Zero
fluid (if used) pressure. Adjust the Actuator/Zero pressure regula-
tor to obtain a minimum pressure of 490 kPag (70 psig / 4.8 barg).
NOTE
NOTE
Содержание Western Research IPS-4
Страница 48: ...2 18 IPS 4 Dual Bench UV IR Analyzer This page intentionally left blank...
Страница 140: ...4 48 IPS 4 Dual Bench UV IR Analyzer This page intentionally left blank...
Страница 172: ...5 32 IPS 4 Dual Bench UV IR Analyzer This page intentionally left blank...
Страница 178: ...6 6 IPS 4 Dual Bench UV IR Analyzer This page intentionally left blank...
Страница 218: ...A 2 IPS 4 Dual Bench UV IR Analyzer Analyzer Light Path UV Optical Bench Schematic...
Страница 219: ...Appendix A Drawings A 3 Analyzer Light Path NDIR Optical Bench Schematic CONDENSING...
Страница 222: ...A 6 IPS 4 Dual Bench UV IR Analyzer Detector Board 100 2046...
Страница 223: ...Appendix A Drawings A 7 Display Interface Board 100 2049...
Страница 225: ...Appendix A Drawings A 9 AC Disconnection Board 100 2077...
Страница 227: ...Appendix A Drawings A 11 Relay Board 100 2050...
Страница 228: ...A 12 IPS 4 Dual Bench UV IR Analyzer Analog Board 100 2047...
Страница 229: ...Appendix A Drawings A 13 MCU Board 100 2045...
Страница 230: ...A 14 IPS 4 Dual Bench UV IR Analyzer Xenon Lamp Power Supply Board 100 2061 UV Optical Bench...
Страница 231: ...Appendix A Drawings A 15 Infrared Source Assembly Interface Board 100 2838 IR Optical Bench...
Страница 232: ...A 16 IPS 4 Dual Bench UV IR Analyzer Electronics Enclosure Wiring Diagram Sheet 1 of 5...
Страница 233: ...Appendix A Drawings A 17 Electronics Enclosure Wiring Diagram Sheet 2 of 5...
Страница 234: ...A 18 IPS 4 Dual Bench UV IR Analyzer Electronics Enclosure Wiring Diagram Sheet 3 of 5...
Страница 235: ...Appendix A Drawings A 19 Disconnect Enclosure Wiring Diagram Sheet 4 of 5...
Страница 236: ...A 20 IPS 4 Dual Bench UV IR Analyzer Disconnect Enclosure Wiring Diagram Sheet 5 of 5...
Страница 238: ...S 2 IPS 4 Dual Bench UV IR Analyzer This page intentionally left blank...