Controller / User Interface
| 4-25
4. If the filter verification routine passes, the analyzer will continue with
normal operations.
If the routine fails, “
Zero Required
” will flash on the
HOME
screen
and will be logged on the
Current Alarms
screen. Also, “
FAILED
” will
be displayed for the
Result
field on the
Display Filter Verification
screen (Figure 4-18b).
Repeat the routine and if the error message does not clear, review
the “Neutral Density Filter Verification Diagnostic Chart“ first, and
perform the
Corrective Action
to try to correct the problem. If the error
message still doesn’t clear, contact AMETEK Service for assistance.
5. To view the results of the filter verification, select
Calibration Control
from the
SETUP
menu and press
Enter
. Scroll down to
Display Filter
Verification
and press
Enter
to display the
Display Filter Verification
screen (Figure 4-18b).
Figure 4-18b.
Display Filter
Verification screen.
Display Filter Verification
Result
PASSED
Date
11/25/2020
Filter Deviation
0.300
Wavelength 228
0.353
►
Wavelength 260
0.333
HOME ALARMS STATUS SETUP
Neutral Density Filter Verification Diagnostic Chart
Alert
Description and Suggested Corrective Action
Filter Verification Fail
This alert can occur if the light intensity has dropped below acceptable
levels.
Corrective Action:
Take appropriate safety precautions to power down the analyzer, open
the Electronics Enclosure, and:
•
Check the glass filter to confirm it is clean and free of scratches. If
not, wipe lightly with a soft material and IPA or acetone. If scratched,
contact AMETEK Service for assistance.
•
Confirm Cell Windows and Mirrors are clean. If not clean them as
described in Chapter 5.
•
If the Cell has an absorbing gas or particulates present, confirm the
Zero gas is clean and pure.
•
Note the light intensity during a Zero gas routine. If low, the lamp may
need to be replaced.
•
If the problem is suspected or found to be caused by Bad Wavelength
Mapping, contact AMETEK Service for assistance.
Содержание Western Research IPS-4
Страница 48: ...2 18 IPS 4 Dual Bench UV IR Analyzer This page intentionally left blank...
Страница 140: ...4 48 IPS 4 Dual Bench UV IR Analyzer This page intentionally left blank...
Страница 172: ...5 32 IPS 4 Dual Bench UV IR Analyzer This page intentionally left blank...
Страница 178: ...6 6 IPS 4 Dual Bench UV IR Analyzer This page intentionally left blank...
Страница 218: ...A 2 IPS 4 Dual Bench UV IR Analyzer Analyzer Light Path UV Optical Bench Schematic...
Страница 219: ...Appendix A Drawings A 3 Analyzer Light Path NDIR Optical Bench Schematic CONDENSING...
Страница 222: ...A 6 IPS 4 Dual Bench UV IR Analyzer Detector Board 100 2046...
Страница 223: ...Appendix A Drawings A 7 Display Interface Board 100 2049...
Страница 225: ...Appendix A Drawings A 9 AC Disconnection Board 100 2077...
Страница 227: ...Appendix A Drawings A 11 Relay Board 100 2050...
Страница 228: ...A 12 IPS 4 Dual Bench UV IR Analyzer Analog Board 100 2047...
Страница 229: ...Appendix A Drawings A 13 MCU Board 100 2045...
Страница 230: ...A 14 IPS 4 Dual Bench UV IR Analyzer Xenon Lamp Power Supply Board 100 2061 UV Optical Bench...
Страница 231: ...Appendix A Drawings A 15 Infrared Source Assembly Interface Board 100 2838 IR Optical Bench...
Страница 232: ...A 16 IPS 4 Dual Bench UV IR Analyzer Electronics Enclosure Wiring Diagram Sheet 1 of 5...
Страница 233: ...Appendix A Drawings A 17 Electronics Enclosure Wiring Diagram Sheet 2 of 5...
Страница 234: ...A 18 IPS 4 Dual Bench UV IR Analyzer Electronics Enclosure Wiring Diagram Sheet 3 of 5...
Страница 235: ...Appendix A Drawings A 19 Disconnect Enclosure Wiring Diagram Sheet 4 of 5...
Страница 236: ...A 20 IPS 4 Dual Bench UV IR Analyzer Disconnect Enclosure Wiring Diagram Sheet 5 of 5...
Страница 238: ...S 2 IPS 4 Dual Bench UV IR Analyzer This page intentionally left blank...