Maintenance and Troubleshooting
| 5-7
Locating a Plug in the Sample System
If analyzer problems occur, they are most likely related to improper
sample system operation, such as plugs or leaks. Plugs in the sample sys-
tem cause problems in analyzer response time, either to changing process
conditions or to Zero fluid. Leaks are potentially dangerous and will even-
tually lead to corrosion problems.
Detecting a Plug in the Sample System
This is best done by observing the analyzer measurement response time
immediately following a Zero function. A typical response is
less than 30
seconds to T90
. Observe your analyzer and learn what a normal response
time is. Use the recorded response time as a reference for detecting the
formation of plugs in the sample system.
If the sample response time is longer than normal, it is most likely due
to plugging somewhere in the sample system. The next step is to locate
and remove the plug. The portion of the sample system with the high-
est risk of plugging is the sample inlet and the portion of the sample inlet
most prone to plugging is the valve on the sample probe or process tap.
One way to confirm that a plug is in the sample inlet is to close the Vent
Line Shut Off valve (in the Oven/Sample System Enclosure) and Zero the
analyzer.
Observe the
Cell Pressure
reading from the
STATUS
screen, and if the
reading approaches the Zero fluid supply pressure, a plug is present
somewhere in the sample inlet.
To avoid damaging the pressure transducer, do not set the pressure
higher than the sample system rating.
After confirming that the plug is somewhere in the sample inlet, take
all appropriate safety precautions and dismantle the sample system,
starting at the Sample Probe or process tap.
Plugs in the vent side of the sample system are rare. Plugs in the vent sys-
tem cause the aspirator drive air to flow back into the Measuring Cell and,
essentially, simulate a
Zero
condition.
!
CAUTION
NOTE
Содержание Western Research IPS-4
Страница 48: ...2 18 IPS 4 Dual Bench UV IR Analyzer This page intentionally left blank...
Страница 140: ...4 48 IPS 4 Dual Bench UV IR Analyzer This page intentionally left blank...
Страница 172: ...5 32 IPS 4 Dual Bench UV IR Analyzer This page intentionally left blank...
Страница 178: ...6 6 IPS 4 Dual Bench UV IR Analyzer This page intentionally left blank...
Страница 218: ...A 2 IPS 4 Dual Bench UV IR Analyzer Analyzer Light Path UV Optical Bench Schematic...
Страница 219: ...Appendix A Drawings A 3 Analyzer Light Path NDIR Optical Bench Schematic CONDENSING...
Страница 222: ...A 6 IPS 4 Dual Bench UV IR Analyzer Detector Board 100 2046...
Страница 223: ...Appendix A Drawings A 7 Display Interface Board 100 2049...
Страница 225: ...Appendix A Drawings A 9 AC Disconnection Board 100 2077...
Страница 227: ...Appendix A Drawings A 11 Relay Board 100 2050...
Страница 228: ...A 12 IPS 4 Dual Bench UV IR Analyzer Analog Board 100 2047...
Страница 229: ...Appendix A Drawings A 13 MCU Board 100 2045...
Страница 230: ...A 14 IPS 4 Dual Bench UV IR Analyzer Xenon Lamp Power Supply Board 100 2061 UV Optical Bench...
Страница 231: ...Appendix A Drawings A 15 Infrared Source Assembly Interface Board 100 2838 IR Optical Bench...
Страница 232: ...A 16 IPS 4 Dual Bench UV IR Analyzer Electronics Enclosure Wiring Diagram Sheet 1 of 5...
Страница 233: ...Appendix A Drawings A 17 Electronics Enclosure Wiring Diagram Sheet 2 of 5...
Страница 234: ...A 18 IPS 4 Dual Bench UV IR Analyzer Electronics Enclosure Wiring Diagram Sheet 3 of 5...
Страница 235: ...Appendix A Drawings A 19 Disconnect Enclosure Wiring Diagram Sheet 4 of 5...
Страница 236: ...A 20 IPS 4 Dual Bench UV IR Analyzer Disconnect Enclosure Wiring Diagram Sheet 5 of 5...
Страница 238: ...S 2 IPS 4 Dual Bench UV IR Analyzer This page intentionally left blank...