Installation and Start-Up
| 3-29
5. After the leak check has been performed and has passed, close the
Bench Heater
fuse and
Oven Heater
fuse (if Oven/Sample System
Enclosure is heated) to apply power to the Measuring Cell and Oven
Heater temperature-control zone circuitry. Ensure that the sample
system door has been closed and latched.
Watch the purge controller until the indicators turn GREEN (left) and
BLACK (right). Install the key in the bypass valve and place it in the
“
ACTIVE
” position. Remove the bypass and enclosure keys and keep
them in a safe place.
No alarms will be displayed on the
HOME
screen or entered into the
Alarm Log
until AFTER the system has left the
Not Ready
state
for the first time after power-up, which occurs after the entire sample
system has stabilized (2–3 hours) and reached normal operating
temperature.
6. While performing the remaining start-up tests, close the analyzer
doors to allow the sample system to warm up to normal operating
temperature (necessary only if Oven/Sample System Enclosure is
heated).
The Oven and the rest of the sample system will reach normal oper-
ating temperature in approximately 2 hours, however (application-
specific) analyzers may take up to another hour to stabilize.
To minimize the time it takes for the entire sample system to warm
up to normal operating temperature, be sure to close the doors after
performing the checks that require them to be open.
NOTE
NOTE
Содержание Western Research IPS-4
Страница 48: ...2 18 IPS 4 Dual Bench UV IR Analyzer This page intentionally left blank...
Страница 140: ...4 48 IPS 4 Dual Bench UV IR Analyzer This page intentionally left blank...
Страница 172: ...5 32 IPS 4 Dual Bench UV IR Analyzer This page intentionally left blank...
Страница 178: ...6 6 IPS 4 Dual Bench UV IR Analyzer This page intentionally left blank...
Страница 218: ...A 2 IPS 4 Dual Bench UV IR Analyzer Analyzer Light Path UV Optical Bench Schematic...
Страница 219: ...Appendix A Drawings A 3 Analyzer Light Path NDIR Optical Bench Schematic CONDENSING...
Страница 222: ...A 6 IPS 4 Dual Bench UV IR Analyzer Detector Board 100 2046...
Страница 223: ...Appendix A Drawings A 7 Display Interface Board 100 2049...
Страница 225: ...Appendix A Drawings A 9 AC Disconnection Board 100 2077...
Страница 227: ...Appendix A Drawings A 11 Relay Board 100 2050...
Страница 228: ...A 12 IPS 4 Dual Bench UV IR Analyzer Analog Board 100 2047...
Страница 229: ...Appendix A Drawings A 13 MCU Board 100 2045...
Страница 230: ...A 14 IPS 4 Dual Bench UV IR Analyzer Xenon Lamp Power Supply Board 100 2061 UV Optical Bench...
Страница 231: ...Appendix A Drawings A 15 Infrared Source Assembly Interface Board 100 2838 IR Optical Bench...
Страница 232: ...A 16 IPS 4 Dual Bench UV IR Analyzer Electronics Enclosure Wiring Diagram Sheet 1 of 5...
Страница 233: ...Appendix A Drawings A 17 Electronics Enclosure Wiring Diagram Sheet 2 of 5...
Страница 234: ...A 18 IPS 4 Dual Bench UV IR Analyzer Electronics Enclosure Wiring Diagram Sheet 3 of 5...
Страница 235: ...Appendix A Drawings A 19 Disconnect Enclosure Wiring Diagram Sheet 4 of 5...
Страница 236: ...A 20 IPS 4 Dual Bench UV IR Analyzer Disconnect Enclosure Wiring Diagram Sheet 5 of 5...
Страница 238: ...S 2 IPS 4 Dual Bench UV IR Analyzer This page intentionally left blank...