Calibration
SG Series Programming
7-16
M550129-03 Rev L
7.
Set the programming source to 0.5V ±1mV.
8.
Verify that the unit is set to 10% ± 0.25% of full-scale output voltage. If
necessary adjust R35 (on the isolated interface adapter) for 10% output voltage.
9.
Repeat the steps above as required to obtain the required accuracy.
10. Remove all connections to the remote analog control connector, J1 (Figure 7-6).
Place the jumper between pins 5 and 6. This reverts the SGI series unit back to
local, front panel control.
Figure 7-6. Analog Control Connector (J1)