SG Series Programming
Calibration
M550129-03 Rev L
7-3
7.3
VOLTAGE PROGRAMMING CALIBRATION
(ETHERNET)
STEP
DESCRIPTION
1. Issue a *CLS command.
2. Issue a *RST command.
3.
Program the overvoltage protection to maximum to prevent nuisance
trips:
CAL:OUTP:VOLT:PROT:DAC 65535
4.
Program the output current to full scale to prevent Constant-Current
operation:
CAL:OUTP:CURR:DAC 65535
5.
Program the output of the first calibration point by sending the following
command string from the computer:
CAL:OUTP:VOLT:DAC 3275
6. Let the output settle and measure the voltage with a high precision
voltmeter; this is value 1.
7.
Enter the actual output voltage value of the first calibration point:
CAL:OUTP:VOLT:FIVEPOINT 1 <value 1>
8.
Program the output of the second calibration point by sending the
following command string from the computer:
CAL:OUTP:VOLT:DAC 19000
9. Let the output settle and measure the voltage with a high precision
voltmeter; this is value 2.
10.
Enter the actual output voltage value of the second calibration point:
CAL:OUTP:VOLT:FIVEPOINT 2 <value 2>
11.
Program the output of the third calibration point by sending the following
command string from the computer:
CAL:OUTP:VOLT:DAC 32000
12. Let the output settle and measure the voltage with a high precision
voltmeter; this is value 3.
Содержание Sorensen SG Series
Страница 2: ......
Страница 4: ...Contents SG Series Programming viii M550129 03 Rev L This page intentionally left blank...
Страница 6: ...Contents SG Series Programming x M550129 03 Rev L This page intentionally left blank...
Страница 8: ...Contents SG Series Programming xii M550129 03 Rev L This page intentionally left blank...
Страница 14: ...Contents SG Series Programming xviii M550129 03 Rev L This page intentionally left blank...
Страница 16: ...Overview SG Series Programming 1 2 M550129 03 Rev L This page intentionally left blank...
Страница 126: ...SCPI Command Operation SG Series Programming 6 60 M550129 03 Rev L This page intentionally left blank...
Страница 143: ...M550129 03 Rev L 8 1 SECTION 8 SCPI STATUS IMPLEMENTATION...
Страница 144: ...SCPI Status Implementation SG Series Programming 8 2 M550129 03 Rev L This page intentionally left blank...
Страница 146: ...This page intentionally left blank...