Calibration
SG Series Programming
7-14
M550129-03 Rev L
7.8.1
Adjustment for Current Mode
STEP
DESCRIPTION
1. Set the SGI Series unit to operate in remote current programming mode using
an external 0-5 VDC voltage source as shown in
+
-
1
13
23 IP RTN
COM 6
+
-
16 IP 10V
IP 5V 10
0-5 VDC
VOLTAGE
SOURCE
0-10 VDC
VOLTAGE
SOURCE
Figure 7-3. Remote Current Programming Using 0-5 VDC or 0-10 VDC Voltage Source
2. Attach a precision meter across the shunt Kelvin terminals.
3. Attach a precision meter in parallel with the voltage programming source.
4. Set the voltage source to 0.0V.
5.
Apply AC power and press the "Output On/Off" switch (
) to the ON
position.
Figure 7-4. SGI Front Panel
6. Set the voltage source to 5.0V ± 1mV.
On/Off Switch
Содержание Sorensen SG Series
Страница 2: ......
Страница 4: ...Contents SG Series Programming viii M550129 03 Rev L This page intentionally left blank...
Страница 6: ...Contents SG Series Programming x M550129 03 Rev L This page intentionally left blank...
Страница 8: ...Contents SG Series Programming xii M550129 03 Rev L This page intentionally left blank...
Страница 14: ...Contents SG Series Programming xviii M550129 03 Rev L This page intentionally left blank...
Страница 16: ...Overview SG Series Programming 1 2 M550129 03 Rev L This page intentionally left blank...
Страница 126: ...SCPI Command Operation SG Series Programming 6 60 M550129 03 Rev L This page intentionally left blank...
Страница 143: ...M550129 03 Rev L 8 1 SECTION 8 SCPI STATUS IMPLEMENTATION...
Страница 144: ...SCPI Status Implementation SG Series Programming 8 2 M550129 03 Rev L This page intentionally left blank...
Страница 146: ...This page intentionally left blank...