Calibration
SG Series Programming
7-10
M550129-03 Rev L
9.
Reset the output current to 0 amps.
CAL:OUTP:CURR:DAC 0
10.
Program the Digital Interface Adapter (DIA) to calculate the current readback
calibration gain and offset values:
CAL:OUTP:CURR:CALC
CAL:MEAS:CURR:CALC
11.
Program the DIA to unlock the non-volatile memory for calibration value
storage:
CAL:UNLOCK “6867”
12.
Program the Digital Interface Adapter (DIA) to store the date:
GPIB
:
CAL:DATE YYYYMMDD
Ethernet:
CAL:MOD:LASTCALDATE MM DD YYYY
CAL:MOD:NEXTCALDATE MM DD YYYY
13.
Program the DIA to store the calibration values in non-volatile memory:
CAL:STORE
Allow 15 seconds for the non-volatile memory to be updated with the
new calibration values.
14.
Program the DIA to lock the non-volatile memory for calibration value
protection
CAL:LOCK
Calibration is complete.
Содержание Sorensen SG Series
Страница 2: ......
Страница 4: ...Contents SG Series Programming viii M550129 03 Rev L This page intentionally left blank...
Страница 6: ...Contents SG Series Programming x M550129 03 Rev L This page intentionally left blank...
Страница 8: ...Contents SG Series Programming xii M550129 03 Rev L This page intentionally left blank...
Страница 14: ...Contents SG Series Programming xviii M550129 03 Rev L This page intentionally left blank...
Страница 16: ...Overview SG Series Programming 1 2 M550129 03 Rev L This page intentionally left blank...
Страница 126: ...SCPI Command Operation SG Series Programming 6 60 M550129 03 Rev L This page intentionally left blank...
Страница 143: ...M550129 03 Rev L 8 1 SECTION 8 SCPI STATUS IMPLEMENTATION...
Страница 144: ...SCPI Status Implementation SG Series Programming 8 2 M550129 03 Rev L This page intentionally left blank...
Страница 146: ...This page intentionally left blank...