Calibration
SG Series Programming
7-8
M550129-03 Rev L
10.
Enter the actual output current of the second calibration point:
CAL:OUTP:CURR:FIVEPOINT 2 <value 2>
11.
Program the output of the third calibration point by sending the following
command string from the computer:
CAL:OUTP:CURR:DAC 32000
12.
Let the output settle and measure the output current with the current shunt
and the high precision voltmeter; this is value 3.
13.
Enter the actual output current of the third calibration point:
CAL:OUTP:CURR:FIVEPOINT 3 <value 3>
14.
Program the output of the fourth calibration point by sending the following
command string from the computer:
CAL:OUTP:CURR:DAC 45000
15.
Let the output settle and measure the output current with the current shunt
and the high precision voltmeter; this is value 4.
16.
Enter the actual output current of the fourth calibration point:
CAL:OUTP:CURR:FIVEPOINT 4 <value 4>
17.
Program the output of the fifth calibration point by sending the following
command string from the computer:
CAL:OUTP:CURR:DAC 62250
18.
Let the output settle and measure the output current with the current shunt
and the high precision voltmeter; this is value 5.
19.
Enter the actual output current of the fifth calibration point:
CAL:OUTP:CURR:FIVEPOINT 5 <value 5>
20.
To review entered data for five
–point current calibration, issue the following
query:
CAL:OUTP:CURR:FIVEPOINT?
21.
Program the supply to unlock the non-volatile memory for calibration value
storage:
CAL:UNLOCK “6867”
22.
Program the supply to store the calibration values in non-volatile memory:
CAL:STORE
Allow 15 seconds for the non-volatile memory to be updated with the
new calibration values.
Содержание Sorensen SG Series
Страница 2: ......
Страница 4: ...Contents SG Series Programming viii M550129 03 Rev L This page intentionally left blank...
Страница 6: ...Contents SG Series Programming x M550129 03 Rev L This page intentionally left blank...
Страница 8: ...Contents SG Series Programming xii M550129 03 Rev L This page intentionally left blank...
Страница 14: ...Contents SG Series Programming xviii M550129 03 Rev L This page intentionally left blank...
Страница 16: ...Overview SG Series Programming 1 2 M550129 03 Rev L This page intentionally left blank...
Страница 126: ...SCPI Command Operation SG Series Programming 6 60 M550129 03 Rev L This page intentionally left blank...
Страница 143: ...M550129 03 Rev L 8 1 SECTION 8 SCPI STATUS IMPLEMENTATION...
Страница 144: ...SCPI Status Implementation SG Series Programming 8 2 M550129 03 Rev L This page intentionally left blank...
Страница 146: ...This page intentionally left blank...