Asterion DC Multioutput Series
Programming Manual
M330517-01 Rev C
13-129
Current Programming Calibration
Preparation:
1. Power down the Unit Under Test (UUT) and remove the input power for safety.
2. Allow 5 minutes for the energy in the output to bleed down to a safe level.
3. Attach the precision
shunt between the power supply’s output terminals.
4. Place the electronic load in Constant Resistance mode.
R = (FSV * 80%) / FSC * (Present DAC value / MAX DAC value)
If the UUT is an ASA Model:
FSV = Maximum rated power / Current rating.
If the UUT is an ASM Model:
FSV = Full scale rated voltage.
5. Attach the voltmeter across the shunt.
6. Apply input power and power up the UUT.
7. After the UUT has initialized, begin the calibration procedure.
STEP
DESCRIPTION
1.
Issue a *CLS command.
2.
Issue a *RST command.
3.
Program the supply to unlock the non-volatile memory for calibration value
storage:
CAL<n>:UNLOCK “6867”
4.
Program the overvoltage protection to maximum to prevent nuisance trips:
CAL<n>:OUTP:VOLT:PROT:DAC 65535
(65535 is the max DAC value).
5.
Program the output voltage to full scale to prevent Constant-Voltage operation:
CAL<n>:OUTP:VOLT:DAC 65535
6.
Program the output of the first calibration point by sending the following
command string from the computer:
CAL<n>:OUTP:CURR:DAC 3275
(3275 is the present DAC value for point1)
7.
Let the output settle and measure the output current with the current shunt and
the high precision voltmeter; this is value 1.
8.
Enter the actual output current of the first calibration point:
CAL<n>:OUTP:CURR:FIVEPOINT 1 <value 1>
9.
Program the output of the second calibration point by sending the following
command string from the computer:
Содержание Sorensen Asterion DC ASM
Страница 1: ...M330517 01 Rev C Asterion DC Multioutput Series Programming Manual...
Страница 6: ...Asterion DC Multioutput Series Programming Manual M330517 01 Rev C v SAFETY SYMBOLS...
Страница 17: ...Asterion DC Multioutput Series Programming Manual M330517 01 Rev C xvi This page intentionally left blank...
Страница 20: ...Asterion DC Multioutput Series Programming Manual M330517 01 Rev C 1 3 This page intentionally left blank...
Страница 24: ...Asterion DC Multioutput Series Programming Manual M330517 01 Rev C 2 7 This page intentionally left blank...
Страница 26: ...Asterion DC Multioutput Series Programming Manual M330517 01 Rev C 3 9 This page intentionally left blank...
Страница 33: ...Asterion DC Multioutput Series Programming Manual 4 16 M330517 01 Rev C This page intentionally left blank...
Страница 38: ...Asterion DC Multioutput Series Programming Manual M330517 01 Rev C 5 21 This page intentionally left blank...
Страница 52: ...Asterion DC Multioutput Series Programming Manual M330517 01 Rev C 6 35 This page intentionally left blank...
Страница 55: ...Asterion DC Multioutput Series Programming Manual 7 38 M330517 01 Rev C This page intentionally left blank...
Страница 63: ...Asterion DC Multioutput Series Programming Manual 8 46 M330517 01 Rev C This page intentionally left blank...
Страница 73: ...Asterion DC Multioutput Series Programming Manual 9 56 M330517 01 Rev C This page intentionally left blank...
Страница 115: ...Asterion DC Multioutput Series Programming Manual 10 98 M330517 01 Rev C This page intentionally left blank...
Страница 119: ...Asterion DC Multioutput Series Programming Manual 11 102 M330517 01 Rev C This page intentionally left blank...
Страница 137: ...Asterion DC Multioutput Series Programming Manual 12 120 M330517 01 Rev C This page intentionally left blank...
Страница 145: ...Asterion DC Multioutput Series Programming Manual 13 128 M330517 01 Rev C CAL n LOCK...
Страница 159: ...Asterion DC Multioutput Series Programming Manual 13 142 M330517 01 Rev C This page intentionally left blank...
Страница 162: ...M330461 01 Rev C 14 145...