Serial Communications
|
A-9
Figure A-3.
Main variable table (3).
0E
current output #1 Function
inT
0 - oxygen
2 - cell Temperature
4 - T/c mv
6 - cell mv
0F
current output #2 Function
inT
same as command letter 0E
12
current output #1 20 mA sP
FloAT
13
current output #1 4 mA sP
FloAT
14
current output #2 20 mA sP
FloAT
15
current output #2 4 mA sP
FloAT
1A
current output #1 Filter
inT
0 - 100
1B
current output #2 Filter
inT
0 - 100
1E
Alarm 3 sP
FloAT
in ppm (see command 6c for %)
1F
Alarm 4 sP
FloAT
in ppm (see command 6d for %)
29
Recovery Time
HEX
XXXX - where 0200 means 2:00
2A
span gas value
FloAT
2B
Zero gas value
FloAT
4d
Pressure
FloAT
4E
operation Temp
FloAT
5d
Alarm 3 Function
inT
0 - o2/Flow
1 - in cal
2 - in verify
5E
Alarm config
HEX
Bit #
7 - Future use
6 - Future use
5 - Alarm #4 low o2
4 - Alarm #4 High o2
3 - Future use
2 - Alarm #3 Hi/lo Flow
1 - Alarm #3 low o2
0 - Alarm #3 High o2
5F
gas counter
HEX
cal gas in use
00 - span
01 - Zero
7F - Thermal gas Recovery
80 - cal Recovery
81 - normal operation
60
cal state
HEX
0 - secondary cal
1 - verify
2 - Thermal cal
3 - normal operation
COMMAND
VARIABLE NAME
FORMAT
DATA FORMAT
LETTER
Содержание Dycor CG1000-RTP
Страница 1: ...Dycor CG1000 RTP User Manual Thermox 150 Freeport Road Pittsburgh PA 15238 PN 90355VE Rev N...
Страница 12: ...xii Dycor CG1000 RTP Oxygen Analyzer This page intentionally left blank...
Страница 18: ...1 6 Dycor CG1000 RTP Oxygen Analyzer This page intentionally left blank...
Страница 34: ...3 14 Dycor CG1000 RTP Oxygen Analyzer This page intentionally left blank...
Страница 96: ...B 4 Dycor CG1000 RTP Oxygen Analyzer This page intenionally left blank...