Installation & Start-Up
|
3-3
Sample Gas Connection
1.
close
Valve 1
and
Valve 2
. set the pressure of P1 on the pressure con-
troller (on the vacuum generator) to
zero
(Figure 3-2).
2.
set the three-way valve on the back of the analyzer to
OFF
(Figure 3-1).
3.
1/8”
tubing
connect the sample gas from
Valve 1
to the sample gas
inlet on the back panel of the analyzer using 1/8” tubing.
This connection is labeled
Gas In
(Figure 3-1).
For the best response time, ensure that the sample line between the analyzer
and sample gas is as short as possible and keep the sample tubing diameter
as small as possible.
4.
1/4”
tubing
connect 1/4” tubing from
Gas Outlet
on the back of the
analyzer to
Valve 2
.
The outlet pressure should not exceed 3 PSIG (207 mil-
libar).
5.
use the
Setup/Display
key on the analyzer front panel to select
flow rate
for one of the display lines
Figure 3-2
sample gas set-up.
NOTE
Keep the inlet pressure as low as possible to prevent
damaging the cell or cell O-ring seal. If the cell temperature
changes too fast because of an excessive flow rate, it can
crack the cell. If the inlet pressure exceeds 3 PSIG, it can
damage the cell O-ring seal.
6.
Ensure that the pressure from the pressure controller is less than 3
Psig.open both
Valve 1
and
Valve 2
.
set the three-way valve on the back of the analyzer to
Gas-In
.
7.
Slowly increase the flow of the sample gas into the analyzer to 150 cc/
min. (200 cc/min. maximum) by slowly increasing the P1 pressure from
the pressure controller. View the flow rate on the display.
Keep the inlet pressure as low as possible.
Содержание Dycor CG1000-RTP
Страница 1: ...Dycor CG1000 RTP User Manual Thermox 150 Freeport Road Pittsburgh PA 15238 PN 90355VE Rev N...
Страница 12: ...xii Dycor CG1000 RTP Oxygen Analyzer This page intentionally left blank...
Страница 18: ...1 6 Dycor CG1000 RTP Oxygen Analyzer This page intentionally left blank...
Страница 34: ...3 14 Dycor CG1000 RTP Oxygen Analyzer This page intentionally left blank...
Страница 96: ...B 4 Dycor CG1000 RTP Oxygen Analyzer This page intenionally left blank...