2-2 | Dycor CG1000-RTP Oxygen Analyzer
alarms
Two (2) independent alarms, each high- or low-selectable. Alarms can
be assigned as oxygen, flow, calibrate, or verify. Set relays to ener-
gize or de-energize on alarm.
contact Rating
0.5A, 30V, 10VA max. noninductive load, AC or DC
Diagnostics
Watchdog timer and service alarms. System test for A/D, RAM,
EEPROM and keypad. Display line 4 reserved for full-text error and
diagnostic messages. Twenty-entry exception log for automatically
detected system events.