4-16 | Dycor CG1000-RTP Oxygen Analyzer
Alarm Configure
Define whether alarms should trigger based on high oxygen, low oxygen,
or combined high/low flow condition.
Alarm 3 is set high or low oxygen or combined high/low flow. Alarm 4
is set to trigger high or low oxygen.
Configuring the alarms
1. From the
Alarm
menu, select
Alarm Configure
.
2. Define Alarm 3 as a high oxygen, low oxygen, or combined high/low
flow and Alarm 4 as high oxygen or low oxygen.
3. Press
Cancel
to exit this menu option.
To create two high oxygen alarms, define both Alarm 3 and Alarm 4 as
high oxygen alarms.
Relay Configure
All Series CG1000-RTP relays are Form A—normally open contacts. You
can use the software to set whether the alarms should close on an alarm
condition (energize) or open on an alarm condition (de-energize).
Configuring the relay
1. From the
Alarm
menu, select
Relay Configure
.
2. Select
De-energize on Alarm
or
Energize on Alarm
.
If you select
Energize on Alarm
, the following appears:
Relays are set to
ENERGIZE on Alarm
If you select
De-Energize on Alarm
, the following appears:
Relays are set to
DE-ENERGIZE on Alarm
NOTE
Содержание Dycor CG1000-RTP
Страница 1: ...Dycor CG1000 RTP User Manual Thermox 150 Freeport Road Pittsburgh PA 15238 PN 90355VE Rev N...
Страница 12: ...xii Dycor CG1000 RTP Oxygen Analyzer This page intentionally left blank...
Страница 18: ...1 6 Dycor CG1000 RTP Oxygen Analyzer This page intentionally left blank...
Страница 34: ...3 14 Dycor CG1000 RTP Oxygen Analyzer This page intentionally left blank...
Страница 96: ...B 4 Dycor CG1000 RTP Oxygen Analyzer This page intenionally left blank...