Calibration
|
5-3
Initiate Cal
Perform a calibration.
If you replaced a cell, or your cell has failed a regular calibration
because of a deterioration, you will need to perform a primary calibra-
tion rather than a regular calibration.
Enter the calibration gas value to match the zero calibration gas cylinder
you are using to calibrate your system.
1. From the
Calibrate Key
menu, select
Initiate Cal
.
Have you entered
the Cal/Vrfy values?
2. Select
Continue Cal
to continue, or
Abort Cal
to cancel the calibration
and enter the values or set up calibration plumbing.
Continue Cal
Abort Calibration
You are prompted to run the O
2
span gas.
Inject Span\Verify Gas
ENTER to continue
CANCEL to Abort
3. Set the 3-way valve on the back of the analyzer to the
CAL
position.
4. Set the 3-way valve that delivers the calibration gases to the
SPAN
position.
5. Press
ENTER
once you have applied the span gas. The span gas dis-
play appears.
O2 xx.x % Set xx.x%
Flow x sccm
Press ENTER to Span
6. Adjust the flow rate to 150 sccm. When the O
2
reading on the far left
of the first display is stable, press
ENTER
.
The reading on the far right of the first display line shows the setpoint
value you entered using the
Cal Gas Values
menu option. This set-
point value should match the cal gas cylinder for the O
2
span gas.
NOTE
NOTE
Содержание Dycor CG1000-RTP
Страница 1: ...Dycor CG1000 RTP User Manual Thermox 150 Freeport Road Pittsburgh PA 15238 PN 90355VE Rev N...
Страница 12: ...xii Dycor CG1000 RTP Oxygen Analyzer This page intentionally left blank...
Страница 18: ...1 6 Dycor CG1000 RTP Oxygen Analyzer This page intentionally left blank...
Страница 34: ...3 14 Dycor CG1000 RTP Oxygen Analyzer This page intentionally left blank...
Страница 96: ...B 4 Dycor CG1000 RTP Oxygen Analyzer This page intenionally left blank...