Error Messages & Troubleshooting
|
6-3
Error Messages
Error messages on the control unit display indicate problems with the op-
eration of the analyzer. Identify the type of problem indicated by the error
message and go to “Diagnostic Checks” at the end of this chapter for help
on how to pinpoint the problem.
Always perform the easiest checks first. Only perform the more dif-
ficult checks if the easiest checks don’t locate the problem.
Cell Is Over Temp
The sensor temperature is 30°C or more over correct operating tem-
perature. The software shuts off the furnace until the temperature
returns to normal operating range.
When the temperature has returned to an acceptable range, the message clears. If there is a problem with the sensor
temperature control system, another error message will appear after this message.
Excessive Cal Error
The Primary Calibration has failed because it cannot correct the sen-
sor temperature to within the software’s allowable limits.
Possible reasons for failure are:
- Calibration gas setup problem
- Incorrect flow rates between calibration gases
- Failed pressure sensor
- Failed cell (primary calibration should be run at this time)
- Leak or plug in plumbing
- Cell lead wires polarity reversed
- Flow sensor not calibrated
- Wrong carrier gas selected
Memory is Corrupted
The control unit’s internal EEPROM memory has been corrupted or a
significantly new version of software has been installed in the control
unit. Under these conditions, the software will reprogram the EE-
PROM with factory defaults. Clear this message by setting or reading
any user-programmable value. This message will not clear by remov-
ing and then restoring power to the unit.
If you see this message, you should reset the analyzer to the values required for your application, including
Setup
Key,
Calibrate
Key,
Alarm
Key and
Analog Range
Key settings. In particular, be sure to re-select the carrier gas carrying
the sample (
Setup
key/
Carrier Gas
) and calibrate the flow sensor.
Span Gas Range
Error
Calibration span gas does not read within the software’s allowable
limits. This error causes calibration of the system to abort.
Possible reasons for failure are:
- Calibration gas setup problem
- Incorrect flow rates between calibration gases
- Calibration gas value incorrectly entered
- Failed cell (primary calibration should be run at this time)
- Leak or plug in plumbing
NOTE
Содержание Dycor CG1000-RTP
Страница 1: ...Dycor CG1000 RTP User Manual Thermox 150 Freeport Road Pittsburgh PA 15238 PN 90355VE Rev N...
Страница 12: ...xii Dycor CG1000 RTP Oxygen Analyzer This page intentionally left blank...
Страница 18: ...1 6 Dycor CG1000 RTP Oxygen Analyzer This page intentionally left blank...
Страница 34: ...3 14 Dycor CG1000 RTP Oxygen Analyzer This page intentionally left blank...
Страница 96: ...B 4 Dycor CG1000 RTP Oxygen Analyzer This page intenionally left blank...