5-2 | Dycor CG1000-RTP Oxygen Analyzer
Calibration
Calibration vs. Verification Operations
Verify
during a verify operation, a calibration gas reading is taken, and the
difference between what the system records for the gas value and
its known value (from the gas cylinder) is recorded.
Internal calibra-
tion parameters are not changed.
verify operations are often used to
check that the instrument is running within required tolerances.
Calibration
during a calibration operation, software algorithms adjust the dif-
ference between the current readings and the known calibration
gas values until they are equal.
Span vs. Zero Gas
This chapter also refers to the terms span and zero calibration gases. The
span gas
is the high calibration gas. The
zero gas
is the low calibration gas.
For verify operations only one gas is used.
Aborting a Calibration
You can cancel a calibration or verification at any time by pressing the
Cancel
key.
NOTE