Installation & Start-Up
|
3-9
RS-232 Communications Connections
The CG1000-RTP communicates with other computers using RS-232 serial
communications protocol. The connections used for RS-232 communica-
tions are on connector J7 and labeled as follows (see Figure 3-1 for location
of J7 on back of CG1000-RTP).
dcd
Pin 1
Rd
Pin 2
Td
Pin 3
dTR
Pin 4
gnd
Pin 5
dsR
Pin 6
RTs
Pin 7
cTs
Pin 8
n/A
Pin 9
Select 115/230 Volt Operation
To change the selected voltage:
• Disconnect the AC line power cord from the analyzer.
• Open the top cover of the power entry module using a small-blade
screwdriver.
• Select the voltage by rotating the fuse holder so the desired voltage
aligns with the window, then snap into place.
If you later want to switch the voltage source for the analyzer, you will
also need to change the fuses in the power entry module:
•
115 VAC - Fuse Rating: 2 amp, 250 VAC - Slow Blow
•
230 VAC - Fuse Rating: 1 amp, 250 VAC - Slow Blow
The fuse sizes are 1/4” x 1 1/4”. Do not use 5 x 20 mm fuses.
If the analyzer is a 230-Volt type, the power cord plug must be
changed to the proper type to match your country and/or electrical
code.
NOTE
Содержание Dycor CG1000-RTP
Страница 1: ...Dycor CG1000 RTP User Manual Thermox 150 Freeport Road Pittsburgh PA 15238 PN 90355VE Rev N...
Страница 12: ...xii Dycor CG1000 RTP Oxygen Analyzer This page intentionally left blank...
Страница 18: ...1 6 Dycor CG1000 RTP Oxygen Analyzer This page intentionally left blank...
Страница 34: ...3 14 Dycor CG1000 RTP Oxygen Analyzer This page intentionally left blank...
Страница 96: ...B 4 Dycor CG1000 RTP Oxygen Analyzer This page intenionally left blank...