Controller / User Interface
|
4-11
For verify operations, select one of the following:
Hold during verify
or
Track during verify
Select one of these options, your choice will be confirmed on the display:
Out #1 will TRACK
during Verify
or
Out #1 will HOLD
during Verify
Set Current Mode Option
Define either a 0-to-20 or 4-to-20 mA analog output range for the selected
analog output port.
This output is dependent on both the range of readings you define (see
the
Set Current Range
section), and on the function you track on the
analog output.
Setting the current mode
1. After selecting an analog output port, from the
Analog Range Key,
select
Set Current Mode
.
2. Select 4-to-20 mA Mode or 0-to-20 mA Mode.
Out # y is set for
4 - 20 ma
or
Out # y is set for
0 - 20 ma